Abstract
Purpose
Current PET radiotracer production models result in facility and operational costs that scale prohibitively with the number of tracers synthesized, particularly those made as a single dose-on-demand. Short of a paradigm shift in the technology and economics of radiotracer production, the impact of PET on precision medicine will be limited. Inexpensive, microfluidic radiochemistry platforms have the potential to significantly reduce costs associated with dose-on-demand production and expand the breadth of PET tracers accessible for molecular imaging.
Procedures
To produce a miniaturized dose-on-demand device for [68Ga]Ga-PSMA-11 production, a microfluidic chip was assembled in polydimethylsiloxane (PDMS), combining all components of tracer production in an integrated, compact, and easily utilized platform. On-chip radionuclide concentration, as well as radionuclide and precursor starting material mixing and reaction were incorporated. The radionuclide was sourced from a standard, commercially available 68Ge/68Ga generator. Optimal reaction conditions were determined, which included precursor concentration (5 μg/mL), temperature (95 °C), and reaction time (1 min).
Results
The total trapping efficiency of combined on-chip SCX and SAX columns was greater than 70 % and could be accomplished in ~ 12 min. Under optimized conditions, [68Ga]Ga-PSMA-11 could be reliably synthesized starting from a complete generator elution (1100 MBq [29.7 mCi]) in ~ 12 min, with an average radiochemical yield of 70 %, radiochemical purity > 99 %, and specific activity > 740 MBq/μg (20 mCi/μg). Quality control testing demonstrated that tracer produced using this platform met or exceeded all typical FDA requirements for human use.
Conclusions
A simple, low-cost, dose-on-demand radiosynthesis strategy, such as the chip presented here, represents an opportunity to reduce the financial barriers associated with PET imaging. While this study focused on a device for [68Ga]Ga-PSMA-11, the technology is also applicable to a wide range of other tracers where low-cost, automated, dose-on-demand production is highly desirable.
Similar content being viewed by others
Change history
22 July 2021
This article was updated to correct the captions to Figures 2 and 3, which were reversed.
02 August 2021
A Correction to this paper has been published: https://doi.org/10.1007/s11307-021-01633-w
References
Ametamey SM, Honer M, Schubiger PA (2008) Molecular imaging with PET. Chem Rev 108:1501–1516
Fortt R, Gee A (2013) Microfluidics: a golden opportunity for positron emission tomography? Future Med Chem 5:241–244
Keng PY, van Dam RM (2015) Digital microfluidics: a new paradigm for radiochemistry. Mol Imaging 12
Rensch C, Jackson A, Lindner S, Salvamoser R, Samper V, Riese S, Bartenstein P, Wängler C, Wängler B (2013) Microfluidics: a groundbreaking technology for PET tracer production? Molecules. 18:7930–7956
Audrain H (2007) Positron emission tomography (PET) and microfluidic devices : a breakthrough on the microscale? Angew Chem Int Ed 46:1772–1775
Elvira KS, Solvas XC, Wootton RCR, DeMello AJ (2013) The past, present and potential for microfluidic reactor technology in chemical synthesis. Nat Chem 5:905–915
Wright BD, Whittenberg J, Desai A et al (2016) Microfluidic preparation of a 89Zr-labeled trastuzumab single-patient dose. J Nucl Med 5:747–752
Zhang X, Liu F, Knapp KA, Nickels ML, Manning HC, Bellan LM (2018) A simple microfluidic platform for rapid and efficient production of the radiotracer [18F]Fallypride. Lab Chip 18:1369–1377
Lee C-Y, Wang W-T, Liu C-C, Fu L-M (2016) Passive mixers in microfluidic systems: a review. Chem Eng 288:146–160
Mueller D, Breeman WAP, Klette I, Gottschaldt M, Odparlik A, Baehre M, Tworowska I, Schultz MK (2016) Radiolabeling of DOTA-like conjugated peptides with generator-produced 68Ga and using NaCl-based cationic elution method. Nat Protoc 11:1057–1066
Mueller D, Klette I, Baum RP, Gottschaldt M, Schultz MK, Breeman WAP (2012) Large-scale routine production of 68Ga using 68Zn-pressed target. J Nucl Med 23:1712–1717
Knapp KA, Nickels ML, Manning HC (2019) The current role of microfluidics in radiofluorination chemistry. Mol Imaging Biol 21:1–13
Ahad A, Easwaroorthy B, Zhang H et al (2018) Cyclotron produced 68Ga for PET imaging of prostate cancer. J Nucl Med 59:610
Wang Y, Shao G, Wu J et al (2018) Preparation of 68Ga-PSMA-11 with a synthesis module for micro PET-CT imaging of PSMA expression during prostate cancer progression. Contrast Media. Mol Imaging 2018:8046541
Funding
Funding of this work is from Vanderbilt Center for Molecular Probes, Vanderbilt University Trans-Institutional Programs (TIPS).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
The authors declare that they have no conflict of interest.
This article does not contain any studies with human participants or animals performed by any of the authors.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Zhang, X., Liu, F., Payne, A.C. et al. High-Yielding Radiosynthesis of [68Ga]Ga-PSMA-11 Using a Low-Cost Microfluidic Device. Mol Imaging Biol 22, 1370–1379 (2020). https://doi.org/10.1007/s11307-020-01515-7
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11307-020-01515-7