Introduction

Ion-selective electrodes (ISEs) are important analytical tools that are widely used in many fields related to our daily life, such as clinical diagnostics, environmental monitoring, agriculture, and food industry [1,2,3,4,5]. Thus, there is a great demand for durable, easy-to-use, and less-frequently calibrated ISEs. Over the years, the performance of ISEs has been significantly improved. The detection limit of polymeric membrane-based ISEs was lowered to ca. 10−11 M by adjusting the internal filling solution in the traditional design [6]. The so-called solid contact ion-selective electrodes (SCISEs) were developed in order to overcome limitations of the internal filling solution, i.e., to obtain more durable sensors that are easy to manufacture, can be miniaturized, and require little maintenance. Among various solid contact materials in SCISEs, conducting polymers proved to be useful as their mixed ion-electron conductivity and high redox (pseudo) capacitance improved the potential stability of the SCISE [5, 7]. The conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) doped with polystyrene sulfonate (PSS) has been extensively used as solid contact in the fabrication of SCISEs for ions, such as K+ [8], Ca2+ [9], and lead Pb2+ [9, 10]. The potential of the conducting polymer itself depends both on its redox state (doping level) and on the ion activity in the contacting solution [11, 12]. Thus, it was shown that the standard potential of SCISEs with PEDOT as solid contact can be shifted by applying either a potential which deviates from the open circuit potential of the electrode or by applying current pulses [13]. These results showed that the redox state of PEDOT can be changed electrochemically even after PEDOT is coated by an ion-selective membrane. Also short circuiting of the SCISE with a conventional large-capacitance reference electrode showed to be a promising way of controlling the standard potential of SCISEs [14]. The reversible reduction/oxidation of the conducting polymer solid contact was also utilized to detect two different analytes with a single SCISE [15].

Traditionally, ISE measurements are carried out using the so-called zero-current potentiometric method, where the equilibrium potential of the ISE, measured vs a reference electrode, is dependent on the ion activity of the primary ion in the sample [5]. In recent years, non-zero current measurements of ISEs have attracted considerable attention [16]. Some of these alternative measurement modes are chronopotentiometry [17, 18], voltammetry [19,20,21,22], and constant potential coulometry [23]. Voltammetric methods were introduced to ISEs by employing thin-layer ion-selective membranes [19]. Amemiya et al. reported the analysis of nanomolar concentrations of perchlorate in drinking water using ion-transfer stripping voltammetry [20]. Bakker’s group presented simultaneous detection of multiple ion activities using a thin-layer ionophore-based membrane and cyclic voltammetry [21, 22].

A new constant potential coulometric method was introduced by our group [23, 24]. In this method, the potential of the SCISE vs the reference electrode is kept constant, so that any potential change at the ion-selective membrane (ISM)–sample interface, due to a change in ion activity in the solution, causes oxidation/reduction of the PEDOT solid contact. The analytical signal is thus a transient current that is measured and integrated to obtain the charge corresponding to the change in the logarithm of ion activity in solution [23, 24]. The method was investigated using PEDOT as solid contact in SCISEs for the detection of monovalent ions, such as K+ [24,25,26], Cl [27] and H+ (pH) [26]. The unique feature of this coulometric method is that the analytical signal, i.e., charge (for a given change in ion activity) is proportional to the redox capacitance of the solid contact, meaning that the signal can be amplified by increasing the thickness (loading) of the solid contact layer [24]. This method has the potential to be used for detecting very small concentration changes, e.g., in clinical applications where the relevant concentration ranges of several ions are very narrow. For example, the concentration range of K+ ion in blood serum is 3.5–5.1 mmol dm−3 [1]. An initial drawback of the coulometric method was a relatively long response time [23, 24]. However, by lowering the bulk resistance of the ion-selective membrane, either by increasing the geometrical surface area of the electrode [25] or by using spin-coated thin-layer ISMs [26, 27], it was possible to significantly shorten the response time of the coulometric signal readout method. Based on a capacitance model used earlier for SCISEs [8], the theory of the coulometric readout was developed for Cl-SCISEs [27]. Furthermore, coulometric SCISEs were evaluated for measurement of pH in seawater [26] and K+ in serum [26, 28].

The measured coulometric signal for SCISEs is based on a current transient involving a change in the redox state (doping level) of the conducting polymer (PEDOT) requiring transport of charge-compensating ions through the ISM and PEDOT layers. This means that the coulometric response may be limited by ion transport through the bulk of the ISM or PEDOT, or by ion transfer at the PEDOT–ISM or ISM–solution interfaces. A theoretical capacitive model based on a series RC circuit gave an excellent explanation of the coulometric response of a Cl-SCISE [27]. However, impedance measurements showed that the coulometric response for K+-SCISEs was limited by diffusion even for thin-layer ISMs [26]. Mikhelson et al. presented recently an extended theoretical model for the amperometric and coulometric response of K+-SCISEs by considering also concentration polarization in the ISM [28].

A further study of the coulometric response and the corresponding charge transfer processes of SCISEs is undertaken in this work to strengthen our understanding of the coulometric transduction mechanism. In this work, the transfer of divalent cations in SCISEs is studied using chronoamperometry, constant potential coulometry, and electrochemical impedance spectroscopy. The electrochemical characteristics of SCISEs with different thicknesses of PEDOT(PSS), as well different thicknesses of ISMs, were studied for lead-selective SCISEs (Pb2+-SCISEs) and calcium-selective SCISEs (Ca2+-SCISEs).

Experimental

Materials

Calcium ionophore IV, lead ionophore IV, valinomycin, potassium tetrakis[3,5-bis-(trifluoromethyl)-phenyl]borate (KTFPB), potassium tetrakis(4-chlorophenyl)borate (KTpClPB), tetradodecylammonium tetrakis(4-chlorophenyl)borate (ETH-500), 2-nitrophenyl octyl ether (o-NPOE), bis(2-ethylhexyl)sebacate (DOS), poly(vinyl chloride) (PVC) of high molecular weight, and tetrahydrofuran (THF, > 99.5%) were selectophore reagents purchased from Fluka. Poly(sodium 4-styrenesulfonate) (NaPSS, Mw ~ 70,000), 3,4-ethylenedioxythiophene (EDOT, 97%), calcium chloride, lead nitrate, potassium chloride, and sodium chloride were purchased from Sigma-Aldrich. Deionized water (ELGA Purelab Ultra; resistivity 18.2 MΩ cm) was used for preparing the aqueous solutions.

Electrode preparation

Glassy carbon (GC) disk electrodes (3-mm diameter) used in this work were manufactured in-house by inserting a 3-mm diameter GC rod (SIGRADUR G, Germany) into a hard PVC cylinder using mechanical pressure. The GC disk electrodes were then extensively polished to smooth, mirror-like finish using the following steps: polishing with abrasive papers of different coarseness, then diamond pastes with particle diameters of 15 μm, 9 μm, 3 μm, and 1 μm, respectively, and finally with 0.3-μm aluminum dioxide paste. Immediately after that, the electrodes were ultra-sonicated for 5 min in ethanol and water, respectively. The aqueous polymerization solution (0.01 M EDOT + 0.1 M NaPSS) was purged with N2 gas for 15 min, and N2 flow was kept above the solution during the polymerization procedure. The conducting polymer PEDOT doped with PSS, i.e. PEDOT(PSS), was electrodeposited on the working electrode by galvanostatic electrochemical polymerization using an Autolab General Purpose Electrochemical System (AUT30.FRA2-Autolab, Eco Chemie, B.V., The Netherlands) connected to a conventional one-compartment three-electrode electrochemical cell. The GC electrode was connected as the working electrode, a GC rod as auxiliary (counter) electrode, and a Metrohm double junction Ag/AgCl/3 M KCl/0.1 M NaPSS was used as the reference electrode. To produce polymerization charge (loading) of 1, 5, and 10 mC, a constant current of 0.014 mA (0.2 mA/cm2) was applied for 71, 355, and 710 s, respectively [8].

After polymerization, all the GC/PEDOT(PSS) electrodes were dried overnight in air before covering them with the ISM. Solid contact ion-selective electrodes (SCISEs) with thin-layer of ISM were prepared by spin-coating the corresponding Pb2+-selective or Ca2+-selective membrane cocktail onto the GC/PEDOT(PSS) electrodes placed vertically in a holder on a rotator. The rotation speed was 1500 rpm, and three drops of the membrane cocktail were added dropwise onto the rotating electrode surface. The volume of each drop was ca. 15 μl. A thin layer of ISM was formed, while the excess of the cocktail escaped the electrode surface due to the centrifugal force. The SCISEs with conventional thick-layer ISM were prepared by dropcasting 50 μl aliquot of the corresponding membrane cocktail to cover the whole GC disk surface and the surrounding flat end of the PVC body. The same Pb2+-selective and Ca2+-selective membrane cocktails were used both for spin-coating and dropcasting techniques. The composition of the Pb2+-selective membrane cocktail in % (w/w) was 1% lead ionophore IV, 0.5% KTpClPB, 65.2% o-NPOE, and 33.3% PVC. These components were dissolved in THF (dry fraction of the membrane cocktail = 15%). The composition of Ca2+-selective membrane cocktail in % (w/w) was 0.8% calcium ionophore IV, 0.45% KTFPB, 65.85% o-NPOE, and 32.9% PVC dissolved in THF (dry fraction of the membrane cocktail = 15%). The membranes were dried overnight followed by an overnight conditioning in 1 mM Pb(NO3)2 or 0.01 M CaCl2 before the measurements. The electrodes were stored in the same solutions between the measurements. For comparison, a K+-SCISE with 1 mC PEDOT(PSS) loading covered with dropcast K+-selective ISM was also used [24, 25]. The composition of the K+-selective membrane cocktail in % (w/w) was 1% valinomycin, 0.5% KTFPB, 1% ETH-500, 65.3% DOS, and 32.2% PVC. These components were dissolved in THF (dry fraction of the membrane cocktail = 15%).

Potentiometric and chronoamperometric measurements

Potentiometric calibrations were performed to verify the proper functioning of the Pb2+- and the Ca2+-selective SCISEs. Both potentiometric and coulometric calibrations were done following the protocol from our previous work [26]. In these measurements, Pb2+- and Ca2+-SCISEs with variable loading of the PEDOT(PSS) (polymerization charge 1, 5, and 10 mC) covered either with spin-coated thin-layer ISM or dropcast thick-layer ISM were studied. Three replicates of Pb2+ and Ca2+-SCISEs were prepared for all the experiments.

The starting solution used in the calibration of Pb2+-SCISEs was 100 ml 1 mM Pb(NO3)2 that was diluted with deionized water, whereas for Ca2+-SCISEs, the calibration started in 100 ml 0.01 M CaCl2 + 0.1 M NaCl (as constant ionic background) that was diluted using 0.1 M NaCl. Removing 34.2 ml of the electrolyte and replacing it with equal amount of deionized water or 0.1 M NaCl result in Δlog aPb2+ or Δlog aCa2+ = 0.18 decades/dilution step. A Metrohm double-junction Ag/AgCl/3 M KCl/1 M LiAc was used as the reference electrode in all potentiometric, chronoamperometric, and coulometric measurements.

Electrochemical impedance spectroscopy

Electrochemical impedance spectroscopy (EIS) measurements were performed at the open circuit potential (OCP) of Pb2+-SCISEs and Ca2+-SCISEs in 0.1 M Pb(NO3)2 and 0.1 M CaCl2, respectively. The Autolab Frequency Response Analyzer System (AUT20.FRA2-Autolab, Eco Chemie, B.V.) connected to a conventional one-compartment three-electrode electrochemical cell was used to carry out the measurements. A Metrohm single-junction Ag/AgCl/3 M KCl was used as reference electrode and a glassy carbon rod as the counter electrode. The sinusoidal excitation signal with an excitation amplitude of 10 mV (RMS) around the OCP was used in the frequency range 100 kHz–10 mHz.

Results and discussion

Results from the potentiometric measurements revealed that the slopes of the linear part in the calibration curves for both Pb2+-SCISEs and Ca2+-SCISEs were ca. 26 ± 1 mV/decade. These values were considered to be sufficiently close to the Nernstian value (29.6 mV/decade) to be used further for chronoamperometric and coulometric measurements in this work.

In analogy with the case of monovalent cation-selective electrodes described earlier [23,24,25,26], the oxidation/reduction of the PEDOT(PSS) solid contact in the case of divalent cation-selective electrodes is expected to follow Eq. 1, taking the Pb2+-SCISE as an example.

$$ 2{\mathrm{PEDOT}}^{+}{\left({\mathrm{PSS}}^{-}\right)}_{\mathrm{film}}+{\mathrm{Pb}}_{\mathrm{ISM}}^{2+}+2{\mathrm{e}}^{-}\leftrightarrow {\left[{\left({\mathrm{PEDOT}}^0\left({\mathrm{PSS}}^{-}\right)\right)}_2{\mathrm{Pb}}^{2+}\right]}_{\mathrm{film}} $$
(1)

In the coulometric method, the potential between the working electrode (SCISE) and the reference electrode is kept constant during the chronoamperometric measurement by using a potentiostat. A decrease in the activity of the primary ion Pb2+, due to the dilution of the sample solution, will cause a decrease in the potential of the ISM–solution interface and thus a shift of the potential of the SCISE in the negative direction vs the reference electrode. As a result, the PEDOT(PSS) solid contact will be oxidized in order to compensate this drop in potential. The oxidation current flowing between the SCISE and the counter electrode is measured and integrated to obtain the cumulated charge (Q) that is expected to be linearly proportional to the change in the logarithm of Pb2+ activity.

The chronoamperometric and coulometric response of the Pb2+-SCISEs and Ca2+-SCISEs, with PEDOT(PSS) as solid contact prepared using 1, 5, and 10 mC polymerization charge and covered with 50 μl dropcast ISM, is shown in Fig. 1. It is seen from Fig. 1 a and e that a higher PEDOT(PSS) loading (higher redox capacitance) in the SCISEs leads to longer response times and also larger cumulated charge values. A thicker solid contact film means that a larger amount of PEDOT(PSS) is involved in the oxidation/reduction process and produces a higher charge, at a given activity change of the primary ion in the solution (Fig. 1 b and f). These results are in good qualitative agreement with the results from our previous work using K+-SCISEs [24].

Fig. 1
figure 1

Chronoamperograms (a, e), cumulated charge Q vs time (b, f), Q vs log aPb2+ or log aCa2+ (c, g), and the slope (Q vs log aPb2+ or log aCa2+) vs the polymerization charge of PEDOT(PSS) solid contact (d, h) of Pb2+-SCISEs and Ca2+-SCISEs with 1, 5, and 10 mC PEDOT(PSS) as solid contact covered with 50 μl dropcast thick ISM

As shown in Fig. 1, the current in the chronoamperograms of the Pb2+-SCISEs and Ca2+-SCISEs returned to the baseline within 5 min for the SCISEs having a 1 mC PEDOT(PSS) solid contact. However, for the SCISEs having 5 and 10 mC PEDOT(PSS) solid contacts, a longer equilibration time would be needed to reach equilibrium after successive dilutions. The cumulated charge vs time for Pb2+-SCISEs is presented in Fig. 1b, and the cumulated charge Q vs log aPb2+ is shown in Fig. 1c. As can be seen, the cumulated charge Q is linearly proportional to log aPb2+ and the obtained Q value is amplified by increasing the thickness (redox capacitance) of the solid contact. However, the slope (Q vs log aPb2+) does not increase linearly with the polymerization charge of PEDOT(PSS) as shown in Fig. 1d, which can be explained by the fact that the current response does not return back to zero after successive dilutions for the 5 and 10 mC PEDOT(PSS) solid contacts (Fig. 1a). The Ca2+-SCISEs (Fig. 1g) show the same trend as the Pb2+-SCISEs, i.e., the cumulated charge Q is linearly proportional to log aCa2+. Similarly, for Ca2+-SCISEs, the slope (Q vs log aCa2+) does not increase linearly with the PEDOT(PSS) loading as shown in the inset of Fig. 1h. Clearly, for these SCISEs with a dropcast thick ISM, the amplification effect can be only partly utilized when limiting the response time to 5 min, which was chosen for practical reasons.

The reproducibility of the coulometric response for Pb2+-SCISEs with 1 mC PEDOT(PSS) solid contact covered with dropcast ISM is shown in Fig. 2. As shown in Fig. 2 a and b, the chronoamperometric signals and the cumulated charge Q are almost identical for the measurements that were carried out with two identical Pb2+-SCISEs (Pb2+-SCISE No.1 and Pb2+-SCISE No.2 in Fig. 2) using the same dilution protocol (Δlog aPb2+ = 0.18 decades/dilution step). A slightly higher cumulated charge vs time response for a repeated measurement (Fig. 2b) can be due to slight increase in the sensitivity (potentiometric slope) of the electrode after the first measurement to low concentrations.

Fig. 2
figure 2

Chronoamperograms (a) and cumulated charge Q vs time (b) recorded for Pb2+-SCISEs with 1 mC PEDOT(PSS) covered with dropcast thick ISM. The starting solution is 1 mM Pb(NO3)2 diluted with deionized water with Δlog aPb2+ = 0.18 decades/dilution step

A comparison of the chronoamperograms of a K+-SCISE and a Pb2+-SCISE recorded under five successive dilution steps with Δlog aK+ and Δlog aPb2+ = 0.18 decades/dilution step is shown in Fig. 3. For both electrodes, the solid contact PEDOT(PSS) was prepared with 1 mC polymerization charge and covered with the relevant dropcast thick ISMs. Theoretically, according to the Nernst equation, the potential change of the monovalent K+-SCISE should be twice that of the divalent Pb2+-SCISE for a given dilution step, here Δlog aK+ and Δlog aPb2+ = 0.18 decades/dilution step. Hence, also the cumulated charge Q vs time is expected to be twice as large for the K+-SCISE compared with that for the Pb2+-SCISE. As can be seen in Fig. 3b, the difference in Q is smaller than expected. This apparent deviation from the theoretical prediction can be due to differences in the redox capacitance of PEDOT(PSS) in the presence of Pb2+ compared with K+ or ion-transport limitations within the SCISEs. As seen in Fig. 3a, the Pb2+-SCISE shows a faster response than the K+-SCISE for a given concentration change. Thus, after the same measurement time (5 min), the charge accumulated by the Pb2+-SCISE is closer to the equilibrium value than that accumulated by the K+-SCISE, which will influence the cumulated charge presented in Fig. 3b.

Fig. 3
figure 3

Chronoamperograms (a) and cumulated charge Q vs time (b) for a K+-SCISE and a Pb2+-SCISE with 1 mC PEDOT(PSS) solid contact covered with dropcast thick ISM. The starting solution for the Pb2+-SCISE is 1 mM Pb(NO3)2 diluted with deionized water, and for the K+-SCISE, it is 0.01 M KCl + 0.1 M NaCl that was diluted with 0.1 M NaCl

When comparing the Q vs log a curves for the Pb2+-SCISE and the Ca2+-SCISE (Fig. 1), it is evident that the Pb2+-SCISE gives a considerably higher charge than the Ca2+-SCISE for the same polymerization charge of the PEDOT(PSS) layer. This indicates that the effective redox capacitance of PEDOT(PSS) is significantly higher in the presence of Pb2+ than in the presence of Ca2+ ions.

Figure 4 shows the effect of the ISM thickness on the chronoamperometric and coulometric response for Pb2+-SCISEs and Ca2+-SCISEs with 1, 5, and 10 mC loading of PEDOT(PSS) solid contact covered either with dropcast thick or spin-coated thin ISM. The dilution was again done with 0.18 Δlog aPb2+ or Δlog aCa2+ steps. For the Pb2+-SCISEs, the chronoamperograms (solid line) and the corresponding cumulated charge vs time (dashed line) during two dilution steps are shown in Fig. 4a–c. The corresponding results for the Ca2+-SCISEs are shown in Fig. 4d–f.

Fig. 4
figure 4

Enlargement of the chronoamperograms (solid line) for Pb2+-SCISEs and Ca2+-SCISEs and the corresponding cumulated charge vs time (dashed line) with 1 mC (a, d), 5 mC (b, e), and 10 mC (c, f) PEDOT(PSS) solid contact covered with spin-coated thin-layer (black) or dropcast (red) ISM, respectively. The primary ion activity (log a) after each dilution step is shown in the figure. The results for the Pb2+-SCISEs recorded in Pb(NO3)2 solutions (without background electrolyte) and for the Ca2+-SCISEs in CaCl2 solutions (with 0.1 M NaCl as constant ionic background)

The chronoamperometric and coulometric results in Fig. 4 show that a thin ISM (lower bulk resistance) gives a faster coulometric response than a thick ISM (higher bulk resistance) for both the Pb2+-SCISEs and the Ca2+-SCISEs. The results in Fig. 4 also show that the coulometric signal is amplified with increasing thickness (increasing redox capacitance) of the PEDOT solid contact. These results are in qualitative agreement with results obtained for monovalent cations [24,25,26].

However, the results shown in Fig. 4 illustrate some unique features of the amperometric and coulometric signal when applied to SCISEs for the detection of divalent cations. A detailed comparison of the Pb2+-SCISEs and Ca2+-SCISEs with 1 mC PEDOT(PSS) as solid contact shows three interesting new attributes. Firstly, the chronoamperometric and coulometric responses for the Pb2+-SCISEs are almost independent of the ISM thickness (Fig. 4a), which is not the case for the Ca2+-SCISEs (Fig. 4d). Secondly, the cumulated charge for the Pb2+-SCISEs (Fig. 4a) is about twice as high compared with the Ca2+-SCISEs (Fig. 4d). Thirdly, the coulometric response is slower for the Pb2+-SCISEs (Fig. 4a) than for the Ca2+-SCISEs (Fig. 4d) in the case of spin-coated thin ISMs. These results together indicate that the coulometric response of the Pb2+-SCISE is mainly limited by the PEDOT(PSS) layer, while the coulometric response of the Ca2+-SCISE is mainly limited by the ISM layer. With increasing thickness of the PEDOT(PSS) layer, the differences in the cumulated charge for the Pb2+-SCISEs and Ca2+-SCISEs become smaller, which shows that the redox capacitance of PEDOT(PSS) can be utilized to a smaller extent in the presence of Pb2+ (compared with Ca2+) when the PEDOT(PSS) layer becomes thicker. An increase in the PEDOT(PSS) thickness (and redox capacitance) means that a larger number of ions need to be transported through the PEDOT(PSS) as well as through the ISM for a given change of ion activity in solution. This may then cause larger concentration polarization in either the PEDOT(PSS) layer or in the ISM layer (or both).

Electrochemical impedance spectra for Pb2+-SCISEs (in 0.1 M Pb(NO3)2) and Ca2+-SCISEs (in 0.1 M CaCl2) with 1 and 10 mC PEDOT(PSS) as solid contact and dropcast (thick) ISMs are shown in Fig. 5a. The corresponding results for spin-coated (thin) ISMs are shown in Fig. 5b. The SCISEs with dropcast ISM show a high-frequency semicircle corresponding to the bulk resistance (450 ± 60 kΩ) and geometric capacitance (27 ± 2 pF) of the ISM (Fig. 5a). This semicircle is hardly visible for SCISEs with spin-coated ISMs due to the low bulk resistance (Fig. 5b). The low-frequency part of the impedance spectra shows that the overall resistance of the Ca2+-SCISEs is about twice that of the Pb2+-SCISEs in the case of dropcast ISMs (Fig. 5a).

Fig. 5
figure 5

Electrochemical impedance spectra recorded for Pb2+-SCISEs and Ca2+-SCISEs with 1 and 10 mC PEDOT(PSS) as solid contact, covered either with dropcast thick-layer ISM (a) or spin-coated thin-layer ISM (b). The measurements were done in 0.1 M Pb(NO3)2 and 0.1 M CaCl2 for Pb2+- and Ca2+-SCISEs, respectively. Dc-potential = open circuit potential, and frequency range = 100 kHz−10 mHz

The imaginary impedance (− Z″) values of the lowest frequency point (10 mHz) was used to estimate the low-frequency (redox) capacitance (CL) of the PEDOT(PSS) layer in the SCISEs. The estimated CL values for the Pb2+-SCISEs and Ca2+-SCISEs are shown in Table 1. The results in Table 1 show that the low-frequency capacitance (CL) for the SCISEs increases with increasing polymerization charge of PEDOT(PSS). Interestingly, CL is independent of the ISM thickness for the Pb2+-SCISE. This shows that the oxidation/reduction of PEDOT(PSS) is not limited by ion transport in the ISM but rather due to charge transport limitations in the PEDOT(PSS) layer of the Pb2+-SCISE. On the contrary, for the Ca2+-SCISE, the value of CL decreases with increasing thickness of the ISM, indicating that in this case, the oxidation/reduction of PEDOT(PSS) is limited by ion transport in the ISM for the Ca2+-SCISE.

Table 1 Low-frequency capacitance (CL) values for Pb2+-SCISEs and Ca2+-SCISEs estimated from the impedance spectra in Fig. 5

These impedance results (Fig. 5; Table 1) are in agreement with the coulometric results for Ca2+-SCISEs and Pb2+-SCISEs in Fig. 4. When comparing the influence of the polymerization charge of PEDOT(PSS) for a given ISM thickness, one can see that CL increases by ca. 4 times when the polymerization charge increases from 1 to 10 mC in the case of Pb2+-SCISEs and Ca2+-SCISEs.

The observation that the coulometric response for the Pb2+-SCISE was limited by ion transport in the PEDOT(PSS) layer is intriguing. One possibility is that interactions between Pb2+ and PEDOT(PSS) retard the transport of Pb2+ in and out of PEDOT(PSS). The affinity between various cations and PSS was reported to decrease in the following order [29]: Ba2+ > Pb2+ > Sr2+ > Ca2+ > Cu2+ > Co2+ > Ni2+ > Mg2+ > H+ > K+ > Na+ > Li+. Based on these results, Pb2+ interacts more strongly with PSS than Ca2+ and K+. Therefore, the transport of Pb2+ in PEDOT(PSS) may be limited due to complexation between Pb2+ and PSS. Another possible explanation is that Pb2+ being a “soft” cation interacts with sulfur atoms in PEDOT. This is supported by earlier results showing significant interactions between the “soft” Ag+ cation and poly(3-octylthiophene) [30].

Conclusions

In this work, the transfer of divalent cations (Pb2+, Ca2+) in solid contact ISEs (SCISEs) was studied by chronoamperometry, constant potential coulometry, and electrochemical impedance spectroscopy. The electrochemical characteristics of lead-selective SCISEs (Pb2+-SCISEs) and calcium-selective SCISEs (Ca2+-SCISEs) were investigated by varying the thickness of the PEDOT(PSS) solid contact layer and the thickness of the ion-selective membrane. Results from coulometric measurements show that the cumulated charge is linearly proportional to the logarithm of the activity of the primary ion in the solution within the range where the potentiometric response of the SCISE is linear. The signal can be amplified by increasing the thickness of PEDOT(PSS) in a similar way as in SCISEs for the detection of monovalent cations studied earlier. The results show further that the coulometric response of the Pb2+-SCISE is limited by charge (ion) transport in PEDOT(PSS), while the response of the Ca2+-SCISEs is limited mainly by ion transport in the ion-selective membrane. The limited transport of Pb2+ in PEDOT(PSS) was tentatively explained by interactions between Pb2+ and PEDOT(PSS). Thus, a decrease in the thickness of the ion-selective membrane will only marginally shorten the response time for coulometric Pb2+-SCISEs when PEDOT(PSS) is used as solid contact.