Direct Immersion Single-Drop Microextraction and Continuous-Flow Microextraction for the Determination of Manganese in Tonic Drinks and Seafood Samples

  • Published:
Food Analytical Methods Aims and scope Submit manuscript

Abstract

Two methods based on direct immersion single-drop microextraction (DI-SDME) and continuous-flow microextraction (CFME) were developed. Both methods were used for the determination of manganese by graphite furnace atomic absorption spectrometry. The reagent 1-(2-pyridylazo)-2-naphthol (PAN) was used as a complexing agent and 1-butyl-3-methylimidazolium hexafluorophosphate [C4MIM][PF6] as an extraction solvent. The extraction efficiencies were compared so that some variables, such as the volume of solvent, time of extraction, stirring speed, and flow rate, were optimized using the univariate method. The limits of detection were 3.9 and 7.9 ng L−1 for DI-SDME and CFME methods, respectively. The enrichment factor for DI-SDME was 18 and for CFME was 15. The accuracy of the method was assessed by analyzing the certified reference material of mussel tissue (SEM 2976), and recovery tests ranged from 92 to 112%. The two categories of LPME are presented as excellent alternatives for the determination of manganese in real samples of tonic drinks and seafood, showing benefits such as low consumption of reagents, ease of operation, and low cost. The DI-SDME-based method is superior to the CFME method if practical aspects are considered, especially for two advantages: less sample volume required and less time to reach equilibrium.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  • Abdelhamid HN (2016) Ionic liquids for mass spectrometry: matrices, separation and microextraction. Trends Anal Chem 77:122–138

    Article  CAS  Google Scholar 

  • Ahmadzadeh S, Rezayi M, Faghih-Mirzaei E, Yoosefian M, Kassim A (2015a) Highly selective detection of titanium (III) in industrial waste water samples using meso-octamethylcalix [4] pyrrole-doped PVC membrane ion-selective electrode. Electrochim Acta 178:580–589

    Article  CAS  Google Scholar 

  • Ahmadzadeh S, Rezayi M, Kassim A, Aghasi M (2015b) Cesium selective polymeric membrane sensor based on p-isopropylcalix [6] arene and its application in environmental samples. RSC Adv 5:39209–39217

    Article  CAS  Google Scholar 

  • Ahmadzadeh S, Yoosefian M, Rezayi M (2019) Comprehensive experimental and theoretical investigations on chromium (III) trace detection in biological and environmental samples using polymeric membrane sensor. Int J Environ Anal Chem:1–16

  • Andruch V, Kocúrová L, Balogh IS, Škrlíková J (2012) Recent advances in coupling single-drop and dispersive liquid–liquid microextraction with UV–vis spectrophotometry and related detection techniques. Microchem J 102:1–10

    Article  CAS  Google Scholar 

  • ATSDR U (2012) Toxicological profile for manganese. US Department of Health and Human Services. Public Service, Atlanta

    Google Scholar 

  • Badakhshan S, Ahmadzadeh S, Mohseni-Bandpei A, Aghasi M, Basiri A (2019) Potentiometric sensor for iron (III) quantitative determination: experimental and computational approaches. BMC Chem 13:131

  • Betteridge D, John D (1973) Pyridylazonaphthols (PANs) and pyridylazophenols (PAPs) as analytical reagents. Part II. Spectrophotometric and solvent-extraction studies of complex formations. Analyst 98:390–411

    Article  CAS  Google Scholar 

  • Betteridge D, Fernando Q, Freiser H (1963) Solvent extraction of certain transition metal ions with 1-(2-pyridylazo)-2-naphthol. A study of complex formation and distribution equilibria. Anal Chem 35:294–298

  • Cao J, Liang P, Liu R (2008) Determination of trace lead in water samples by continuous flow microextraction combined with graphite furnace atomic absorption spectrometry. J Hazard Mater 152:910–914

    Article  CAS  Google Scholar 

  • Corsini A, Yih IM, Fernando Q, Freiser H (1962) Potentiometric investigation of the metal complexes of 1-(2-pyridylazo)-2-naphthol and 4-(2-pyridylazo) resorcinol. Anal Chem 34:1090–1093

    Article  CAS  Google Scholar 

  • de la Calle I, Pena-Pereira F, Lavilla I, Bendicho C (2016) Liquid-phase microextraction combined with graphite furnace atomic absorption spectrometry: A review. Anal Chim Acta 936:12–39

    Article  Google Scholar 

  • Donaldson EM, Inman WR (1966) Determination of manganese in high-purity niobium tantalum molybdenum and tungsten metals with PAN. Talanta 13:489–497

  • Fang Z, Dong L, Xu S (1992) Critical evaluation of the efficiency and synergistic effects of flow injection techniques for sensitivity enhancement in flame atomic absorption spectrometry. J Anal At Spectrom 7:293–299

    Article  CAS  Google Scholar 

  • Farajzadeh MA, Nouri N, Khorram P (2014) Derivatization and microextraction methods for determination of organic compounds by gas chromatography. Trends Anal Chem 55:14–23

  • Farzadbeh N, Vardini M, Sheikhloie H (2016) Trace determination of manganese(II) by temperature-controlled/assisted ionic liquid-based dispersive liquid-phase microextraction and electrothermal atomic absorption spectrometry. J Iran Chem Soc 13:715–722

    Article  CAS  Google Scholar 

  • Fouladgar M, Ahmadzadeh S (2016) Application of a nanostructured sensor based on NiO nanoparticles modified carbon paste electrode for determination of methyldopa in the presence of folic acid. Appl Surf Sci 379:150–155

    Article  CAS  Google Scholar 

  • Hou L, Lee HK (2002) Application of static and dynamic liquid-phase microextraction in the determination of polycyclic aromatic hydrocarbons. J Chromatogr A 976:377–385

    Article  CAS  Google Scholar 

  • Jain A, Verma KK (2011) Recent advances in applications of single-drop microextraction: A review. Anal Chim Acta 706:37–65

    Article  CAS  Google Scholar 

  • Jeannot MA, Cantwell FF (1996) Solvent microextraction into a single drop. Anal Chem 68:2236–2240

    Article  CAS  Google Scholar 

  • Jeannot MA, Cantwell FF (1997) Mass transfer characteristics of solvent extraction into a single drop at the tip of a syringe needle. Anal Chem 69:235–239

    Article  CAS  Google Scholar 

  • Kokosa JM (2015) Recent trends in using single-drop microextraction and related techniques in green analytical methods. Trends Anal Chem 71:194–204

    Article  CAS  Google Scholar 

  • Lemos VA, Vieira US (2013) Single-drop microextraction for the determination of manganese in seafood and water samples. Microchim Acta 180:501–507

    Article  CAS  Google Scholar 

  • Lemos VA, Oliveira RV, dos Santos WNL, Menezes RM, Santos LB, Ferreira SLC (2019) Liquid phase microextraction associated with flow injection systems for the spectrometric determination of trace elements. Trends Anal Chem 110:357–366

  • Liu H, Dasgupta PK (1996) Analytical chemistry in a drop. Solvent extraction in a microdrop. Anal Chem 68:1817–1821

    Article  CAS  Google Scholar 

  • Manzoori JL, Amjadi M, Abulhassani J (2009a) Ionic liquid-based single drop microextraction combined with electrothermal atomic absorption spectrometry for the determination of manganese in water samples. Talanta 77:1539–1544

    Article  CAS  Google Scholar 

  • Manzoori JL, Amjadi M, Abulhassani J (2009b) Ultra-trace determination of lead in water and food samples by using ionic liquid-based single drop microextraction-electrothermal atomic absorption spectrometry. Anal Chim Acta 644:48–52

    Article  CAS  Google Scholar 

  • Marcinkowski L, Pena-Pereira F, Kloskowski A, Namiesnik J (2015) Opportunities and shortcomings of ionic liquids in single-drop microextraction. Trends Anal Chem 72:153–168

    Article  CAS  Google Scholar 

  • Marczenko Z, Balcerzak M (2000) Separation, preconcentration and spectrophotometry in inorganic analysis. Elsevier

  • Moinfar S, Khayatian G (2017) Continuous sample drop flow-based microextraction combined with graphite furnace atomic absorption spectrometry for determination of cadmium. Microchem J 132:293–298

    Article  CAS  Google Scholar 

  • Ouyang G, Zhao W, Pawliszyn J (2007) Automation and optimization of liquid-phase microextraction by gas chromatography. J Chromatogr A 1138:47–54

    Article  CAS  Google Scholar 

  • Pena-Pereira F, Lavilla I, Bendicho C (2010) Liquid-phase microextraction approaches combined with atomic detection: a critical review. Anal Chim Acta 669:1–16

    Article  CAS  Google Scholar 

  • Santamaria AB (2008) Manganese exposure, essentiality & toxicity. Indian J Med Res 128:484

  • Santamaria AB, Sulsky SI (2010) Risk assessment of an essential element: manganese. J Toxicol Environ Health Part A 73:128–155

    Article  CAS  Google Scholar 

  • Shemirani F, Behgozin S (2018) Combination of dispersive liquid-liquid microextraction and flame atomic absorption spectrometry for simultaneous preconcentration and determination of manganese and nickel in water and food samples. J Iran Chem Soc 15:1907–1912

    Article  CAS  Google Scholar 

  • Soltani H, Pardakhty A, Ahmadzadeh S (2016) Determination of hydroquinone in food and pharmaceutical samples using a voltammetric based sensor employing NiO nanoparticle and ionic liquids. J Mol Liq 219:63–67

    Article  CAS  Google Scholar 

  • Spietelun A, Marcinkowski Ł, de la Guardia M, Namieśnik J (2014) Green aspects, developments and perspectives of liquid phase microextraction techniques. Talanta 119:34–45

    Article  CAS  Google Scholar 

  • Stanisz E, Werner J, Zgoła-Grześkowiak A (2014) Liquid-phase microextraction techniques based on ionic liquids for preconcentration and determination of metals. Trends Anal Chem 61:54–66

    Article  CAS  Google Scholar 

  • Wei G, Li Y, Wang X (2008) Comparison of efficiencies between single-drop microextraction and continuous-flow microextraction for the determination of methomyl in natural waters. Int J Environ Anal Chem 88:397–408

    Article  CAS  Google Scholar 

  • Ye C, Zhou Q, Wang X (2007) Improved single-drop microextraction for high sensitive analysis. J Chromatogr A 1139:7–13

    Article  CAS  Google Scholar 

Download references

Funding

This study was financed in part by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior - Brasil (CAPES) - Finance Code 001. The authors also acknowledge the financial support of the Conselho Nacional de Desenvolvimento Científico e Tecnológico (311419/2018-6) and the Fundação de Amparo à Pesquisa do Estado da Bahia.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Valfredo Azevedo Lemos.

Ethics declarations

Conflict of Interest

Leane Santos Nunes declares that she has no conflict of interest. Maria das Graças Andrade Korn declares the she has no conflict of interest. Valfredo Azevedo Lemos declares that he has no conflict of interest.

Ethical Approval

Not applicable.

Informed Consent

Not applicable.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nunes, L.S., das Graças Andrade Korn, M. & Lemos, V.A. Direct Immersion Single-Drop Microextraction and Continuous-Flow Microextraction for the Determination of Manganese in Tonic Drinks and Seafood Samples. Food Anal. Methods 13, 1681–1689 (2020). https://doi.org/10.1007/s12161-020-01794-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12161-020-01794-4

Keywords

Navigation