Abstract
In this article, in calibration process of strain cylinder, which is performed according to ISO 376 standard, effect of loading apparatus with different dimensions on measurement result has been investigated. In addition, relative repeatability error of calibration on measurement results, temperature fluctuation, accuracy error of measurements, creep error, as well as the stability and metrological characteristics of indicator have been examined. As a result, it has been determined that use of loading pads with different diameters has significantly affected metrological performance of strain cylinder.
Similar content being viewed by others
References
BS EN 12390-4:2000, Testing Hardened Concrete, Part 4: Compressive Strength—Specification for Testing Machines.
EN ISO 376:2011, Metallic Materials-Calibration of Force-proving Instruments Used for the Static Verification of Uniaxial Testing Machines.
P. Averlant, Strain Cylinders for Compression Testing Machines, IMEKO TC3 Cairo, Egypt, 2005.
J. Cui, R. Kumme and H. Kahmann, Proposal and Investigation of a New Method for Calibration of Strain Cylinders Using Laser Interferometry, IMEKO TC3 Helsinki Finland, 2017.
GTM Measurement Amplifier LT Digitizer Technical Data Sheet, in English, 02/03/2015.
B. Aydemir, H. Dizdar and C. Vatan, An Investigation and Calculation of Uncertainty in Calibration of Force-proving Instruments according to EN ISO 376 Standard, Force Laboratory Technical Document, TUBITAK UME, S.2, N.1, 2019 (1–11).
B. Aydemir, H. Dizdar and C. Vatan, General force metrology training notes, TUBITAK UME Gebze-Kocaeli, 2017.
B. Aydemir, Calibration of concrete test press training notes, TUBITAK UME Gebze-Kocaeli, 2011.
EURAMET cg-04, Guide document, uncertainty of force measurements, 2011.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Dİzdar, H., Aydemİr, B. & Vatan, C. Investigation of Different Compression Apparatus Usage Effects in Calibration Process of Strain Cylinders. MAPAN 35, 359–364 (2020). https://doi.org/10.1007/s12647-020-00375-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12647-020-00375-4