Article

Nanostructured MOS Sensor for the Detection, Follow up, and Threshold Pursuing of Campylobacter Jejuni Development in Milk Samples

by
Estefanía Núñez-Carmona
1,*,
Marco Abbatangelo
2,
Dario Zappa
2,
Elisabetta Comini
2,3,
Giorgio Sberveglieri
2,3 and
Veronica Sberveglieri
1,3
1
Institute of Bioscience and Bioresources, CNR-IBBR, 50019 Sesto Fiorentino (FI), Italy
2
Department of Information Engineering, University of Brescia, 25123 Brescia (BS), Italy
3
NANO SENSOR SYSTEMS, Dep. Information Engineering, NASYS spin-off University of Brescia, 25123 Brescia (BS), Italy
*
Author to whom correspondence should be addressed.
Sensors 2020, 20(7), 2009; https://doi.org/10.3390/s20072009
Submission received: 7 February 2020 / Revised: 23 March 2020 / Accepted: 2 April 2020 / Published: 3 April 2020

Abstract

:
Food poisoning is still the first cause of hospitalization worldwide and the most common microbial agent, Campylobacter jejuni, is the most commonly reported gastrointestinal disease in humans in the EU (European Union) as is reported by the European Union One Health 2018 Zoonoses Report styled by the EFSA (European Food Safety Authority) and ECDC (European Center for Disease Prevention and Control). One of the vehicles of transmission of this disease is milk. Nanostructured MOS (Metal Oxide Semiconductor) sensors have extensively demonstrated their ability to reveal the presence and follow the development of microbial species. The main objective of this work was to find a set up for the detection and development follow up of C. jejuni in milk samples. The work was structured in two different studies, the first one was a feasibility survey and the second one was to follow up the development of the bacteria inside milk samples. The obtained results of the first study demonstrate the ability of the sensor array to differentiate the contaminated samples from the control ones. Thanks to the second study, it has been possible to find the limit of microbial safety of the contaminated milk samples.

1. Introduction

Organizations worldwide run and rule all aspects of the food safety, security, and trading at a global level. Organizations such as the EFSA (European Food Safety Agency), WHO (World Health Organization), and FAO (Food and Agriculture Organization) guarantee the most safe food industry environment we have ever enjoyed, putting the risk perception regarding the food industry at the bottom of the concerning list of those in non-developing countries [1,2]. Even with all the measurements applied to reach food safety worldwide, food poisoning is still the first cause of hospitalization globally. The CDC (Center of Disease Control) estimates that each year, 48 million people get sick from a foodborne illness, 128,000 are hospitalized, and 3000 die just in the USA alone [3].
The European Union One Health 2018 Zoonoses Report styled by the EFSA and ECDC (European Food Safety Authority and European Centre for Disease Prevention and Control) approved 19 November, 2019 encloses a list of the Zoonoses included in compulsory annual monitoring (Directive 2003/99/EC List A) [4].
This report claims that Campylobacteriosis is the most commonly reported gastrointestinal disease in humans in the EU and has been so since 2005. The most common sources for the FBOs (food borne outbreaks) were milk and broiler meat, as in previous years [4,5]. The etiologic vector of the illness is Campylobacter jejuni, which is associated with warm-blooded animals involving livestock animals that have shown the microorganisms in their feces [6,7]. Some strains of C. jejuni produce a thermolabile enterotoxin (CJT), [8,9] that affects the gastrointestinal system.
Campylobacter spp. infection can also be transmitted through unpasteurized milk ingestion when a cow has C. jejuni cells in its udder or when milk is contaminated with manure. Campylobacter jejuni is a common raw milk contaminant and is notoriously difficult to isolate from food products, because of its fastidious growth requirements [10,11,12,13,14].
Recently, different kinds of sensor have been used for the detection of C. jejuni. Some involve the use of DNA detection or oligonucleotide hybridization with the specific target of the pathogen using optical [15], acoustic [16], and electrochemical [17,18] techniques. In other cases, quartz crystal microbalance (QCM) hybridized immunosensors using monoclonal and polyclonal antibodies have been used [19]. Normally DNA or immune based techniques are costly and require a high level expertise to be applied. In recent research, two-dimensional nanostructured materials have been widely used for gas sensing [20]. Furthermore, previous research [21] have used MOS nanostructured gas sensors for the detection of the aforementioned pathogen using lab standard conditions. The main differences regarding the volatile compounds present in the head space between the contaminated samples and the controls ones (performed as sterile liquid culture media) reside in the presence of a higher concentration in alcohol compounds in C. jejuni samples than the control one. This increased concentration of alcohols, together with the neoformation of pyrazines, comes as a result of sugar fermentation.
Due to the proven possibility of C. jejuni detection using MOS gas sensors, a rapid, easy to use, and economic alarm system that provides enough and accurate information could be constructed in order to actuate corrective and timely actions in order to prevent contamination and spread of the pathogen in the dairy industry. In this work, we illustrate how through the use of nanostructured MOS sensors it is possible to detect, follow up, and conduct threshold detection pursuing of Campylobacter jejuni development in milk samples with a fast and economic approach. The use of this kind of technology have been largely applied to obtain remarkable results in the agri food field [22,23,24,25,26].

2. Materials and Methods

All the research described in this paper was performed in two different sets of analysis including different techniques. The first one corresponds to a feasibility study of the detection of the pathogen in milk samples, the second one was dedicated to reducing the initial concentration of the microbial agent in the challenge test in order to define the threshold detection of the S3.

2.1. Samples Preparation

Starter cultures were prepared in the same way for the two sets of analysis following the same conditions used in previous work [21] with some modifications. Campylobacter jejuni subsp. jejuni type strain cells were purchased from DMSZ (The Leibniz Institute DSMZ—German Collection of Microorganisms and Cell Cultures GmbH-Deutsche Sammlung von Mikroorganismen und Zellkulturen-), DSM number 4688, (ATCC 33560, CCUG 11284, CIP 702, NCTC 11351). Microbial cells were inoculated in tubes containing 9 mL of Brain Heart Infusion Broth (BHI) liquid media purchased from Sigma Aldrich (Merck). Tubes were incubated in aerobic conditions for 24 h at 35 °C. After the incubation, the culture was used to inoculate tubes of sterile BHI media in order to perform the subsequent dilution.
For the first set of analysis, the same optical density (OD) as of the number 3 standard of McFarland that corresponds to a concentration of 9 × 108 CFU (Colony Forming Unit) by mL [27] was reached. Afterward, sterile BHI media tubes were used to perform serial dilutions in order to decrease two orders of magnitude reaching 9 × 106 CFU/mL. Subsequently sterile chromatography vials of 20 mL containing 4 mL of sterile whole milk were inoculated with 100 μL using the previous solution (9 × 106 CFU/mL), which leads to a final concentration in the vial near 2.20 × 105 CFU/mL. Control samples were performed using 20 mL sterile chromatographic vials containing 4 mL of sterile whole milk.
Regarding the second set of analysis, the samples were performed in the same way as the previous, but reducing the microbial loading in the final samples by an order of magnitude, which leads to a final concentration in a vial of 2.20 × 104 CFU/mL. Control samples were performed exactly in the same way as the first set of analysis. For this set, the analysis time was performed in just 7 h, so a total number of nine samples was prepared from which two were control samples.

2.2. Microbial Analysis

Plate count technique has been applied in every step of the different analysis for the first set of analysis, so every 21 h of analysis from 0 h to 93 h in order to verify, first the number of inoculated cells and second, how the microorganisms will develop inside the sample. The used media were BHI agar plates that once inoculated were incubated at 35 °C for 48 h before counting. Regarding the second set of analysis of measurement, plate count was applied at 0 h and at the end of the analysis after 7 h following the same conditions.

2.3. S3 Analysis

In order to operate the system, the S3 instrument was coupled with the autosampler headspace system HT2010H (HTA s.r.l., Brescia, Italy). This autosampler is equipped with a thermostatic shaking oven to equilibrate the sample headspace and a 42-loading position carrousel. The S3 device and the sensor arrays used in this work were constructed and designed by the SENSOR Laboratory (University of Brescia, Italy) in collaboration with NASYS S.r.l., a spin-off of the University of Brescia. This device has demonstrated remarkable performances as has been broadly reported in the literature [28,29,30,31]. S3 device components can be divided in three different groups:
  • Pneumatic components: that moves the gas sample to the sensing chamber.
  • Electronic boards: that are in charge of data acquisition and transmission to the dedicated WebApp and at the same time allows the synchronization between S3 and the HT2010H.
  • Sensing chamber in which up to 10 MOX sensor could be located and is thermally isolated from the surrounding environment.
The sensing array is composed of different sensing elements based on metal oxide semiconducting (MOS) materials. Each sensor consists of a MOS sensing layer deposited on an insulating alumina substrate. The sensing principle is the variation of the electrical conductance of the sensing material upon interaction with the surrounding atmosphere. Target molecules react with adsorbed oxygen species on sensing the material surface, releasing electrons that modulate the electrical properties including the electrical conductance.
For this purpose, two interdigited platinum electrodes were deposited by sputtering technique on top of the sensing material. The kinetic of the chemical reaction was thermally promoted. Therefore, on the backside of the substrate, we deposited a platinum heater to precisely control the working temperature of each device. Finally, each substrate was mounted on a gold-coated “transistor outline” (TO-39) package by electrosoldered gold wires (Figure 1).
To perform the base line of the sensors, the surrounding atmosphere was filtered using an aluminum cylinder of 21.5 cm in length and 5 cm in width, filled with active carbons. Two different arrays were used for this analysis, one for the first set and the other for the second set of analysis using similar sensors in order to find the best performing combination.
To construct the sensor array of the S3 device, six MOX gas sensors were selected for each step of the analysis. Both used arrays were a mix between tin oxide nanowire [32,33] sensors grown by means of the vapor liquid solid technique [34], and the other sensors were prepared with RGTO (Rheotaxial Growth and Thermal Oxidation) thin film technology [35]. The sensor array of the first and second set of analysis are described and represented in Table 1, which shows the material, the sensor type, the functionalization (if this is the case), and the operating temperature. Since the sensors are aspecific, the selection of sensors with different materials and architecture were selected in order to cover all the volatile finger prints of the different samples. For each sensor, the response (AR/R) to 5 ppm of ethanol, selectivity (response ethanol/response carbon monoxide), and limit of detection (LOD) of ethanol for both S3 sensors arrays have been represented (Table 1).
The analysis carried out by the S3 system for each sample can be divided in different steps for a total analysis time of 30 min by vial:
  • 5 min of sample gas in contact with the sensors array.
  • 5 min to clean the actuators (tubes, pump, sensing chamber, etc.) of the gas sample.
  • 20 min to restore the base line.
The sample arrangements for the two different sets of analysis was different. Regarding the first one (feasibility study), analysis was carried out from 0 to 93 h with a total number of 168 samples that were analyzed continuously. Since the used auto sampler has 42 loading positions, four steps of 21 hours were performed in sequence in order to have continuous control of the changes in the head space of the samples. To the extent of also having a continuous survey of the control samples, seven of the aforementioned samples were positioned inside every carrousel in order to perform them every 3 h at a rate of 5 (150 min) contaminated samples to one control (30 min).
In the second set (define the threshold), the total analysis time was performed in only one step, of 7 h of analysis, with a total number of samples of nine. Control samples were analyzed every 3 h with a rate of five to one as well as in the first set of analysis.
In previous work, C. jejuni was detected in standard laboratory conditions using the culture media as background for the detection of C. jejuni volatiloma [21]. Since the sensors are aspecific and the ultimate goal is to obtain an array that is able to detect the presence of the pathogen in real conditions, new training with real samples (milk) became mandatory to train the sensor array.

2.4. S3 Data Analysis

Al the sensors responses were recorded in terms of resistance (Ω) and at the same time normalized to the first value of the acquisition (R0) using MATLAB® R2015a software (MathWorks, USA). Regarding the feature extraction, ΔR/R0 was used for the data analysis. It was performed calculating the difference between the first value and the minimum value for all sensors. Once the data matrix was calculated, principal component analysis (PCA) was applied to the data in order to verify the increment of the CFU/mL in relationship with the variation of the VOCs (Volatile Organic Compound).

3. Results and Discussion

In the first set of analysis, the theoretical initial concentration of microbial cell inoculation in whole milk at T0 was 2.20 × 105 CFU/mL. Count plate technique confirmed the real number of inoculated cells was 2.92 × 105 CFU/mL at T0, while the control samples where completely clean throughout the entire analysis time of the first set of analysis. The conversely inoculated sample exhibited an increase in the cell number during the analysis time of the first set of analysis, arriving at 2.19 × 109 CFU/mL, enhancing the initial concentration of nearly four orders of magnitude.
Moreover, regarding the results obtained with the S3 device in the first set of analysis, data are represented in the form of PCA in Figure 2.
A clear separation of two clusters is evident in Figure 1, diversifying contaminated samples out of contaminated ones. Apart from these data, no evolution of the contaminated samples was evident which formed a cluster with no differentiation between the less contaminated and the ones that exhibited a big contamination at the end of the 93 h. This output is mainly due to a saturation phenomenon of the sensor response.
In the second set of analysis, a reduction of one order of magnitude was performed in order to find out if it was possible to follow up the microbial development of C. jejuni inside whole milk samples. The results obtained with the plate count technique were in accordance with those obtained with the sensor device S3.
Regarding the sensor response, the obtained results are represented in Figure 3 as a form of PCA. The PCA score plot showed the first two principal components (PCs). The total explained variance (EV) was equal to 91.08% (60.78% for PC1 and 30.3% for PC2), meaning that most of the information was contained in two variables. A clear distinction between the contaminated samples and control milk was achieved considering the samples’ scores along PC1.
Focusing attention on the contaminated specimens, the scores along PC2 can be assessed, which is helpful to understand the point where the threshold of 105 is passed. Indeed, bearing in mind that a concentration of 1.5 × 104 has been calculated from plate count at t = 0 h and C. jejuni doubles each 85 minutes [36], sample S5 contained 8.32 × 104 CFU/mL, while sample S6 was 1.06 × 105 CFU/mL, which oversteps the microbial acceptability for raw products that correspond to 105 CFU/mL. In light of this result, the contaminated raw milk is not suitable for human consumption.

4. Conclusions

The obtained results pave the way for the S3 device to be applied in the dairy industry to prevent contaminations by C. jejuni. This paper illustrates the reaching of two different objectives, the first one was the differentiation between contaminated and not contaminated milk samples, and the second one was the follow up of the microbial development of the microbial specie and the limit detection of the milk salubrity. The obtained results completely fulfill the initial aim of the work. As future work, research will focus on the limit of detection of the instrument trying to reach the minimum detectable concentration in order to be able to actuate corrective actions before the milk reaches the not suitable for human consumption state. In addition, effort will concentrate into study investigating the effect of adding normal milk microflora to the samples in order to observe the interaction between the microorganisms and the volatiloma emitted from them. Experiments adding non-related microorganisms in order to demonstrate the specificity of detection is already on-going.

Author Contributions

Conceptualization, E.N.-C. and V.S.; Methodology, E.N.-C.; Software, M.A.; Validation, E.N.-C., M.A., and V.S.; Formal analysis, E.N.-C. and M.A.; Writing—original draft preparation, E.N.-C., M.A., D.Z. and E.C. Writing—review and editing, V.S. and G.S.; Supervision, G.S., V.S., E.C. and D.Z. All authors have read and agreed to the published version of the manuscript.

Funding

This work was partially funded by MIUR “Smart Cities and Communities and Social Innovation” project titled “SWaRM Net/Smart Water Resource Management-Networks”.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Yeung, R.M.W.; Morris, J. Food safety risk—Consumer perception and purchase behavior. Br. Food J. 2001, 103, 170–187. [Google Scholar] [CrossRef]
  2. Yu, H.; Neal, J.A.; Sirsat, S.A. Consumers’ food safety risk perceptions and willingness to pay for fresh-cut produce with lower risk of foodborne illness. Food Cont. 2018, 86, 83–89. [Google Scholar] [CrossRef]
  3. Centers for Disease Control and Prevention, Foodborne Illnesses and Germs. Available online: https://www.cdc.gov/foodsafety/foodborne-germs.html (accessed on 21 December 2019).
  4. European Food Safety Authority and European Centre for Disease Prevention and Control (EFSA and ECDC). The European Union One Health 2018 Zoonoses Report; Scientific Report; European Food Safety Authority: Parma, Italy, 2019. [CrossRef] [Green Version]
  5. Artursson, K.; Schelin, J.; Lambertz, S.T.; Hansson, I.; Engvall, E.O. Foodborne pathogens in unpasteurized milk in Sweden. Int. J. Food Microbiol. 2018, 284, 120–127. [Google Scholar] [CrossRef] [PubMed]
  6. Açik, M.N.; Çetinkaya, B. Heterogeneity of Campylobacter jejuni and Campylobacter coli strains from healthy sheep. Vet. Microbiol. 2006, 115, 370–375. [Google Scholar] [CrossRef]
  7. Brown, P.E.; Christensen, O.F.; Clough, H.E.; Diggle, P.J.; Hart, C.A.; Hazel, S.; Kemp, R.; Leatherbarrow, A.J.H.; Moore, A.; Sutherst, J.; et al. Frequency and spatial distribution of environmental Campylobacter spp. Appl. Environ. Microbiol. 2004, 70, 6501–6511. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. Klipstein, F.A.; Engert, R.F. Properties of Crude Campylobacter jejuni Heat-Labile Enterotoxin. Infect. Immun. 1984, 45, 314–319. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  9. Jay, J.M.; Loessner, M.J.; Golden, D.A. Foodborne Gastroenteritis Caused by Vibrio, Yersinia and Campylobacter. In Modern Food Microbiology, 7th ed.; Heldman, D.R., Ed.; Heldman Associates: San Marcos, CA, USA, 2005; Volume 1, pp. 657–678. ISBN 0-387-23180-3. [Google Scholar]
  10. Boysen, L.; Vigre, H.; Rosenquist, H. Seasonal influence on the prevalence of thermotolerant Campylobacter in retail broiler meat in Denmark. Food Microbiol. 2011, 28, 1028–1032. [Google Scholar] [CrossRef] [PubMed]
  11. Humphrey, T.J.; Beckett, P. Campylobacter jejuni in dairy cows and raw milk. Epidemiol. Infect. 1987, 98, 263–269. [Google Scholar] [CrossRef] [Green Version]
  12. Davis, K.R.; Dunn, A.C.; Burnett, C.; McCullough, L.; Dimond, M.; Wagner, J.; Smith, L.; Carter, A.; Willardson, S.; Nakashima, A.K. Campylobacter jejuni Infections Associated with Raw Milk Consumption—Utah, 2014. In Morbidity and Mortality Weekly Report; CDC: Atlanta, GA, USA, 2016; Volume 65, pp. 301–305. [Google Scholar] [CrossRef]
  13. Lovett, J.; Francis, D.W.; Hunt, J.M. Isolation of Campylobacter jejuni from raw milk. Appl. Environ. Microbiol. 1983, 46, 459–462. [Google Scholar] [CrossRef] [Green Version]
  14. Del Collo, L.P.; Karns, J.K.; Biswas, D.; Lombard, E.J.; Haley, B.J.; Kristensen, R.C.; Kopral, C.A.; Fossler, C.P.; Van Kessel, J.A.S. Campy e milk Prevalence, antimicrobial resistance, and molecular characterization of Campylobacter spp. in bulk tank milk and milk filters from US dairies. J. Dairy Sci. 2017, 100, 3470–3479. [Google Scholar] [CrossRef] [Green Version]
  15. Cecchini, F.; Manzano, M.; Mandabi, Y.; Perelman, E.; Marks, R.S. Chemiluminescent DNA optical fibre sensor for Brettanomyces bruxellensis detection. J. Biotechnol. 2012, 157, 25–30. [Google Scholar] [CrossRef] [PubMed]
  16. Jia, K.; Toury, T.; Ionescu, R.E. Fabrication of an atrazine acoustic immunosensor based on a drop-deposition procedure. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2012, 59, 2015–2021. [Google Scholar] [PubMed]
  17. Kubicˇárová, T.; Fojta, M.; Vidic, J.; Tomschik, M.; Suznjevic, D.; Palecˇek, E. Voltammetric and chronopotentiometric measurements with nucleic acid-modified mercury film on a glassy carbon electrode. Electroanal. 2000, 12, 1390–1396. [Google Scholar] [CrossRef]
  18. Vidic, J.; Manzano, M.; Chang, C.M.; Jaffrezic-Renault, N. Advanced biosensors for detection of pathogens related to livestock and poultry. Vet. Res. 2017, 48, 1–22. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  19. Masdor, N.A.; Altintas, Z.; Tothilla, I.E. Sensitive detection of Campylobacter jejuni using nanoparticles enhanced QCM sensor. Biosens. Bioelectron. 2016, 78, 328–336. [Google Scholar] [CrossRef] [PubMed]
  20. Liu, X.; Ma, T.; Pinna, N.; Zhang, J. Two-Dimensional Nanostructured Materials for Gas Sensing. Adv. Funct. Mater. 2017, 27, 1702168. [Google Scholar] [CrossRef]
  21. Núñez-Carmona, E.; Abbatangelo, M.; Sberveglieri, V. Innovative Sensor Approach to Follow Campylobacter jejuni Development. Biosens. 2019, 9, 8. [Google Scholar] [CrossRef] [Green Version]
  22. Sberveglieri, V.; Núñez-Carmona, E.; Pulvirenti, A. Detection of microorganism in water and different food matrix by electronic nose. In Smart Sensors, Measurement and Instrumentation; Springer: Cham, Switzerland, 2015; Volume 11, pp. 243–258. [Google Scholar]
  23. Sberveglieri, V.; Bhandari, M.P.; Núñez-Carmona, E.; Betto, G.; Sberveglieri, G. A novel MOS nanowire gas sensor device (S3) and GC-MS-based approach for the characterization of grated Parmigiano Reggiano cheese. Biosensors 2016, 6, 60. [Google Scholar] [CrossRef] [Green Version]
  24. Núñez-Carmona, E.; Sberveglieri, V.; Ponzoni, A.; Zappa, D.; Pulvirenti, A. Detection of microbial contamination in potable water by Nanowire technology. In Proceedings of the International Conference on Sensing Technology, Liverpool, UK, 2–4 September 2014; pp. 136–139. [Google Scholar]
  25. Núñez-Carmona, E.; Abbatangelo, M.; Zottele, I.; Piccoli, P.; Tamanini, A.; Comini, E.; Sberveglieri, G.; Sberveglieri, V. Nanomaterial gas sensors for online monitoring system of fruit jams. Foods 2019, 8, 632. [Google Scholar] [CrossRef] [Green Version]
  26. Abbatangelo, M.; Núñez-Carmona, E.; Duina, G.; Sberveglieri, V. Multidisciplinary approach to characterizing the fingerprint of Italian EVoO. Molecules 2019, 24, 1457. [Google Scholar] [CrossRef] [Green Version]
  27. McFarland Standards. Available online: http://www.dalynn.com/dyn/ck_assets/files/tech/TM53.pdf (accessed on 20 January 2020).
  28. Abbatangelo, M.; Núñez-Carmona, E.; Sberveglieri, V.; Zappa, D.; Comini, E.; Sberveglieri, G. Application of a Novel S3 Nanowire Gas Sensor Device in Parallel with GC-MS for the Identification of Rind Percentage of Grated Parmigiano Reggiano. Sensors 2018, 18, 1617. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  29. Abbatangelo, M.; Núñez-Carmona, E.; Sberveglieri, V. Application of a novel S3 nanowire gas sensor device in parallel with GC-MS for the identification of Parmigiano Reggiano from US and European competitors. J. Food Eng. 2018, 236, 36–43. [Google Scholar] [CrossRef]
  30. Sberveglieri, V. Validation of Parmigiano Reggiano Cheese Aroma Authenticity, Categorized through the Use of an Array of Semiconductors Nanowire Device (S3). Materials 2016, 9, 81. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  31. Sberveglieri, V.; Bhandari, M.P.; Nunez Carmona, E.; Betto, G.; Soprani, M.; Malla, R.; Sberveglieri, G. Spectrocolorimetry and nanowire gas sensor device S3 for the analysis of Parmigiano Reggiano cheese ripening. In Proceedings of the 2017 ISOCS/IEEE International Symposium on Olfaction and Electronic Nose (ISOEN), Montreal, QC, Canada, 28–31 May 2017. [Google Scholar]
  32. Ponzoni, A.; Zappa, D.; Comini, E.; Sberveglieri, V.; Faglia, G.; Sberveglieri, G. Metal oxide nanowire gas sensors: Application of conductometric and surface ionization architectures. Chem. Eng. Trans. 2012, 30, 31–36. [Google Scholar]
  33. Sberveglieri, G.; Concina, I.; Comini, E.; Falasconi, M.; Ferroni, M.; Sberveglieri, V. Synthesis and integration of tin oxide nanowires into an electronic nose. Vacuum 2012, 86, 532–535. [Google Scholar] [CrossRef]
  34. Wagner, R.S.; Ellis, W.C. Vapor-Liquid-Solid Mechanism of Single Crystal Growth. Appl. Phys. Lett. 1964, 4, 89–90. [Google Scholar] [CrossRef]
  35. Sberveglieri, G. Recent developments in semiconducting thin-film gas sensors. Sens. Actuators B-Chem. 1995, 23, 103–109. [Google Scholar] [CrossRef]
  36. Battersby, T.; Walsh, D.; Whyte, P.; Bolton, D.J. Campylobacter growth rates in four different matrices: Broiler caecal material, live birds, Bolton broth, and brain heart infusion broth. Infect. Ecol. Epidemiol. 2016, 6. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Scheme of the sensor architecture. From top to the bottom: two interdigited platinum electrodes (gray), the sensing material (dark gray), alumina substrate (yellow), platinum heather (gray), gold-coated “transistor outline” (TO-39) package (brown).
Figure 1. Scheme of the sensor architecture. From top to the bottom: two interdigited platinum electrodes (gray), the sensing material (dark gray), alumina substrate (yellow), platinum heather (gray), gold-coated “transistor outline” (TO-39) package (brown).
Sensors 20 02009 g001
Figure 2. Principle component analysis (PCA) score plot representing the data obtained during the 93 h of the first set of analysis where contaminated samples are dots colored in green while control samples are yellow squares. The horizontal axis represents the explained variance enclosed in PC1 as equal to 60.31%, conversely, the vertical axis represents the explained variance enclosed in PC2 as equal to 23.04%. The total explained variance enclosed in the graphic is equal to 83.35%.
Figure 2. Principle component analysis (PCA) score plot representing the data obtained during the 93 h of the first set of analysis where contaminated samples are dots colored in green while control samples are yellow squares. The horizontal axis represents the explained variance enclosed in PC1 as equal to 60.31%, conversely, the vertical axis represents the explained variance enclosed in PC2 as equal to 23.04%. The total explained variance enclosed in the graphic is equal to 83.35%.
Sensors 20 02009 g002
Figure 3. PCA score plot representing the data obtained during the first 7 h of the second set of analysis. The horizontal axis represents the explained variance enclosed in PC1 as equal to 60.78%, conversely, the vertical axis represents the explained variance enclosed in PC2 as equal to 30.30%. The total explained variance enclosed in the graphic is equal to 91.08%.
Figure 3. PCA score plot representing the data obtained during the first 7 h of the second set of analysis. The horizontal axis represents the explained variance enclosed in PC1 as equal to 60.78%, conversely, the vertical axis represents the explained variance enclosed in PC2 as equal to 30.30%. The total explained variance enclosed in the graphic is equal to 91.08%.
Sensors 20 02009 g003
Table 1. Type, composition, morphology, operating temperature, response (ΔR/R), selectivity (response ethanol/response carbon monoxide), and limit of detection (LOD) of ethanol for both S3 sensors arrays made at the SENSOR Laboratory.
Table 1. Type, composition, morphology, operating temperature, response (ΔR/R), selectivity (response ethanol/response carbon monoxide), and limit of detection (LOD) of ethanol for both S3 sensors arrays made at the SENSOR Laboratory.
Materials (Type)CompositionMorphologyOperating Temperature (°C)Response to 5 ppm of EthanolSelectivityLimit of Detection (LOD) of Ethanol (ppm)
First Step Array
SnO2Au (n)SnO2 functionalized with Au clustersRGTO400 °C6.530.5
SnO2 (n)SnO2RGTO300 °C3.52.51
SnO2 (n)SnO2RGTO400 °C420.8
SnO2Au+Au (n)SnO2 grown with Au and functionalized with gold clustersNanowire350 °C72.50.5
SnO2Au (n)SnO2 grown with AuNanowire350 °C52.11
CuO (p)CuONanowire400 °C1.51.51
Second Step Array
SnO2Au (n)SnO2 grown with AuRGTO400 °C6.530.5
SnO2 (n)SnO2RGTO400 °C420.8
CuO (p)CuONanowire350 °C1.51.51
SnO2Au (n)SnO2 grown with AuNanowire350 °C52.11
SnO2 (n)SnO2Nanowire350 °C420.8

Share and Cite

MDPI and ACS Style

Núñez-Carmona, E.; Abbatangelo, M.; Zappa, D.; Comini, E.; Sberveglieri, G.; Sberveglieri, V. Nanostructured MOS Sensor for the Detection, Follow up, and Threshold Pursuing of Campylobacter Jejuni Development in Milk Samples. Sensors 2020, 20, 2009. https://doi.org/10.3390/s20072009

AMA Style

Núñez-Carmona E, Abbatangelo M, Zappa D, Comini E, Sberveglieri G, Sberveglieri V. Nanostructured MOS Sensor for the Detection, Follow up, and Threshold Pursuing of Campylobacter Jejuni Development in Milk Samples. Sensors. 2020; 20(7):2009. https://doi.org/10.3390/s20072009

Chicago/Turabian Style

Núñez-Carmona, Estefanía, Marco Abbatangelo, Dario Zappa, Elisabetta Comini, Giorgio Sberveglieri, and Veronica Sberveglieri. 2020. "Nanostructured MOS Sensor for the Detection, Follow up, and Threshold Pursuing of Campylobacter Jejuni Development in Milk Samples" Sensors 20, no. 7: 2009. https://doi.org/10.3390/s20072009

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop