Early-stage cervical cancer diagnosis based on an ultra-sensitive electrochemical DNA nanobiosensor for HPV-18 detection in real samples

Abstract

Background

In several years ago, infection with human papillomaviruses (HPVs), have been prevalent in the worlds especially HPV type 18, can lead to cervical cancer. Therefore, rapid, accurate, and early diagnosis of HPV for successful treatment is essential. The present study describes the development of a selective and sensitive electrochemical biosensor base on DNA, for early detection of HPV-18. For this purpose, a nanocomposite of reduced graphene oxide (rGO) and multiwalled carbon nanotubes (MWCNTs) were electrodeposited on a screen-printed carbon electrode (SPCE). Then, Au nanoparticles (AuNPs) were dropped on a modified SPCE. Subsequently, single strand DNA (ssDNA) probe was immobilized on the modified electrode. The link attached between AuNPs and probe ssDNA provided by l-cysteine via functionalizing AuNPs (Cys-AuNPs). The differential pulse voltammetry (DPV) assay was also used to electrochemical measurement. The measurement was based on the oxidation signals of anthraquninone-2-sulfonic acid monohydrate sodium salt (AQMS) before and after hybridization between the probe and target DNA.

Results

The calibration curve showed a linear range between 0.01 fM to 0.01 nM with a limit of detection 0.05 fM. The results showed that the optimum concentration for DNA probe was 5 µM. The good performance of the proposed biosensor was achieved through hybridization of DNA probe-modified SPCE with extracted DNA from clinical samples.

Conclusions

According to the investigated results, this biosensor can be introduced as a proprietary, accurate, sensitive, and rapid diagnostic method of HPV 18 in the polymerase chain reaction (PCR) of real samples.

Introduction

Several studies demonstrated that HPV is commonly spread by sexual activity. HPVs are classified according to their risk of cancer infection into two groups (high risk and low-risk). Among the high-risk HPVs, the most dangerous types are HPV-16 and HPV-18 that are responsible for almost more than half of cervical cancers. Also, the high-risk HPVs are responsible for 36% of penile cancers, 51% of vulvar cancer, 63% of oropharyngeal carcinomas, 64% of vaginal cancers, 93% of anal cancers and 96% of cervical cancers [1]. Double-stranded DNA genomes of HPVs encode eight genes, which among them, E6 and E7 by having transforming properties, are necessary for malignant conversion [2]. Traditional methods for detection of HPVs are Pap smear and visual inspection with acetic acid. These methods have their own limitations such as poor specificity and sensitivity [3]. However, due to recent advances in molecular methods, this method is used to detect and also identify the types of HPV. For example, COBAS®4800, digene HC2 High-Risk HPV DNA, linear array®, PapilloCheck®, INNO-LiPA, clinical arrays®HPV, CLART®human papillomavirus 2 and others are the assays that showed a high sensitivity for the diagnosis of HPVs and pre-cancerous lesions. Nevertheless, the methods mentioned have limitations such as high cost, long response time, requirment of trained personnel as well as advanced equipment and the need for PCR products [4, 5]. Therefore, considering many disadvantages of the existing diagnostic methods, the development of an appropriate diagnostic method for identifying HPV is essential. In recent years, DNA biosensors are introduced as a new appropriate device for rapid, simple, economical, sensitive, specific and early detection of pathogens [5,6,7]. They are diagnostic devices that are designed for detection of DNA target among million-fold excess of non-target species, by hybridized complementary DNA probes immobilized on the surface of various types of electrodes [8]. Until now, different biosensors for HPVs assays were reported, but electrochemical biosensors have more advantageous due to their portability, cost-effectiveness, small sample volume, small size, and ease of use [9,10,11]. In order to improve the performance of the biosensor, various materials are used to modify the surface of the electrode. Research conducted in recent years shows that rGO is a suitable candidate for improving the performance of electrochemical activities in biosensors. The structure of rGO honeycomb produces significant either chemical or physical properties, such as high surface area, high electrical conductivity and high chemical stability. Furthermore, the introduction of MWCNTs into the matrix of rGO-based materials not only improves electron transfer but also provides more porous structure, which is a very effective way to increase sensitivity and improve biosensor performance [12,13,14]. Moreover, recent advances in biosensors indicate that the use of Au nanoparticles increases electrical conductivity and provides the immobilization passages of DNA on the electrode surface [15]. Many of the amino acids have special properties that make it possible to use as an effective ingredient to immobilize biomolecules on the surface of the electrode. They have special groups that are capable of binding with biomolecules. This improves the immobilization of DNA [16]. The amino acid l-Cysteine is an effective ingredient for electrode surface modification. It has a thiol group, which can bond strongly to Ag or Au electrodes through bonding of amino groups to DNA [17,18,19]. Due to the high prevalence of HPV and the lack of appropriate diagnostic methods in the primary stages of infection, in this investigation a sensitive and accurate diagnostic method for the specific identification of HPV 18 is introduced. It is well known that the development of the novel sensing nanomaterials such as rGO, MWCNT and AuNPs has been proven as an effective method to fabricate electrochemical DNA biosensors for diagnosis of cancer markers. To the best of our knowledge, this is the first report of functionalized electrochemical DNA-biosensor for HPV-18 detection based on rGO-MWCNT/l-Cys-AuNPs nanocomposite in the real samples.

Materials and method

Reagents and materials

All chemicals and reagents were of analytical reagent grade. All aqueous solutions were prepared using deionized water with resistivity of ∼ 18 MΩ cm. Anthraquninone-2-sulfonic acid monohydrate sodium salt (AQMS), l-cysteine and sodium chloride were purchased from Merck (Germany). Multi-walled carbon Nano tube (MWCNT), gold nanoparticles (AuNPs) and sulfuric acid were purchased from Sigma Aldrich (USA). The potassium permanganate and graphite flakes were purchased from R & M Chemicals (U.K). The applied solvents throughout the experiments such as ethanol, sulphuric acid, phosphoric acid, hydrochloric acid, and hydrogen peroxide were in analytical grade and purchased from Merck (Germany). All single-stranded DNA (ssDNA) was purchased from TAG Copenhagen Company (Denmark). Screen-printed carbon electrodes were supplied by DropSens (Asturias).

Apparatus

The electrochemical measurement was carried out with a Galvanostat–Potentiostat; (Autolab PG-STAT-204). The voltammetry measurements were performed on a three-electrode system. The working electrode was a screen printed carbon electrode (4 mm OD). Ag/AgCl, 3 M KCl, RE-5B (7.5 cm × 6 mm OD) and platinum wire electrode (BASi®, USA) were used as a reference and auxiliary electrodes, respectively. Electrochemical impedance spectroscopy (EIS) analysis was conducted using a Galvanostat-Potentiostat, Autolab 302 N with FRA2 impedance module controlled by Nova 1.11 software, Metrohm, Switzerland. The ultrasonic homogenizer (Sonopuls, HD 3200) was applied throughout the experiments for homogenizing and breakdown of nanoparticles. All hybridization experiments were carried out in a water bath (Memmert, WNB 14).

Human papillomavirus synthetic DNA oligonucleotides

All oligonucleotide stock solutions were diluted with phosphate-buffer (0.05 M, pH = 7.0) and kept frozen at − 20 Â°C when not used. The following three oligonucleotide sequences were used in this study (Table 1).

Table 1 The oligonucleotide-based DNA sequences used in the current study

DNA extraction, PCR method and HPV genotyping

For extracting DNA from endocervical of patients infected with HPV 18, the High Pure Viral Nucleic Acid Kit (Roche Life Science, Mannheim Germany) was used. For HPV testing, a DNA-based liquid-crystal display (LCD)-array kit (Chipron GmbH, Berlin, Germany) was applied as described previously [20, 21]. The PCR program is presented in Table 2. In total, the PCR was accomplished in Veriti® 96-Well Thermal Cycler (Thermo Fisher Scientific; Applied Biosystems, USA).

Table 2 The PCR cycling conditions

The positive control had DNA from a HPV 18 case while the negative control had no DNA. According to the manufacturer’s protocol (Chipron GmbH) PCR, hybridization, labeling, and staining were carried out. The slide was scanned by SlideReader Software v2.0 (Chipron GmbH), followed by gel electrophoresis.

Preparation of the modified SPCE

For this purpose, DI water was used for washing the bare SPCE and then the electrode was electrochemical polished using cyclic voltammetry (CV) in 0.1 M H2SO4, ranging from (0.5 to 1.5 V) with a scan rate of 50 mVs−1. The modified Hummer’s method was applied for the synthesis of GO based on our previous work. Prior to deposition, the solution of GO and MWCNT was prepared by dispersing 5.0 mg of GO and 11.0 mg of MWCNT into 0.2 M PBS, pH 6.5 after 2 h of sonication. Electrodeposition of rGO-MWCNT nanocomposite on a SPCE in a one-step approach was performed by CV method through scanning between (− 1.0 and 0 V) at a scan rate of 50 mVs−1 until 10 cycles (Fig. 1a).

Fig. 1
figure 1

Cyclic voltammograms of electrodeposition of (a) rGO-MWCNT nanocompoite on a bare SPCE (in 0.2 M PBS, pH 6.5 and Scan rate: 50 mVs−1), b of l-cysteine (10 mM) on a modified SPCE (in 0.1 M PBS, pH 7.4; Scan rate: 50 mVs−1), c the effect of different scan rates on electrodeposition CV curves of rGO-MWCNT nanocompoite in 1 × 10−3 M [Fe(CN)6]−3/[Fe(CN)6]−4 (1:1) containing 0.1 M KCl solution and d Raman spectra of GO, GO with MWCNT and rGO-MWCNT nanocompoite

The AuNP colloidal solution was made by dissolving 1 mg of the powder in 300 Âµl mixture of DI water and ethanol (1:1) followed by 1 h sonication and then drop cast onto the surface of SPCE. l-cysteine (10 mM) in 0.1 M PBS, pH 7.4 was electrochemically deposited on the modified SPCE through CV scanning between (−0.2 and 1.5 V); 10 cycles; scan rate of 50 mV s−1 (Fig. 1b).

As illustrated in Fig. 1a, MWCNT was deposited on the SPCE in the potential range of − 0.6 to − 0.8 V. Electrodeposition of GO on SPCE in the potential range from − 0.8 to − 1.0 V led to slightly reduce oxygen functional groups which resulted to increase the access to carbon groups and its efficiency. Also, according to Fig. 1a, it can be seen that in with increasing the number of cycles, the cathodic peak current increased due to the increment in the amount of rGO-MWCNTs deposited on the surface of SPCE during each cycle, and enhance the electrical conductivity [12, 22, 23].

Figure 1b shows the CV curves of electrodeposition of l-cysteine in the potential range of − 0.2 V to + 1.5 V with a scan rate of 50 mV s−1. A reduction peak occurs at 0.43 V during the reduction scan and related to the SH group. The reduction peak current increased with the increment of the cycle number. The reaction indicated in Scheme 1, occurs during the electrodeposition process [24,25,26]. The electrodeposition of rGO-MWCNT on SPCE at variable scan rates revealed a capacitive improvement with decreasing the rate of scans from 100 to 50 mV/S (see Fig. 1c; 314 < 316 < 324 ÂµF). The capacitance was calculated based on the following equations [27]:

Scheme 1
scheme 1

Immobilization mechanism for l-cysteine onto SPCE

$$C^{\prime} = {I \mathord{\left/ {\vphantom {I {(\Delta V/\Delta t)}}} \right. \kern-0pt} {(\Delta V/\Delta t)}}$$
(1)
$$C = {{C^{\prime}} \mathord{\left/ {\vphantom {{C^{\prime}} {Area}}} \right. \kern-0pt} {Area}}\,or\,{{C^{\prime}} \mathord{\left/ {\vphantom {{C^{\prime}} {mass}}} \right. \kern-0pt} {mass}}$$
(2)

where, I and ∆V/∆t are the measured current and voltage scan rate, respectively.

As illustrated, the electrodeposition of rGO-MWCNT with a scan rate of 50 mV/S on SPCE has a large integrated CV area indicating of higher capacitance (324 ÂµF). Figure 1d report the Raman spectra of synthesized GO, GO with MWCNT and rGO-MWCNT nanocomposite which shows the prominent peaks of D band at ~ 1358 cm−1 and G band at ~ 1587 cm−1, respectively. It is well known that, the D band arising from structural disorder, while G band arising from in-plane C–C bond stretches [27,28,29,30]. The 2D mode is also contribute to the double resonance of sp2 carbon domains and occurs at 2550 to 2800 cm−1. As seen in Fig. 1d the existence of MWCNT demonstrated a broad 2D peak. The 2D band in the Raman spectrum of the rGO-MWCNT becomes more intense, indicating that the graphene sheets are well separated due to reducing the carboxylic groups after electrodeposition process. The intensity ratio of D band to G band (ID/IG) is taken as a measure of disorder. It is obvious from Fig. 1d that the ID/IG for the rGO-MWCNT nanocomposite is lower than GO-MWCNT which indicates the disordered carbon structure. The ratio of 2D to D + G in GO-MWCNT and rGO-MWCNT is higher than GO due to its broad 2D peak. The Raman spectroscopy results revealed that the resulting morphology well incorporates both rGO and MWCNT in the nanocomposite during the electrodeposition process.

Immobilization of HPV-18 DNA probe

In order to immobilize ssDNA probe on the modified SPCE by the dip-coating method, the electrode was placed in the solution of HPV-18 DNA (1 ÂµM) in 0.05 M PBS, pH 7.0 for 24 h at 25 Â°C. Then, to remove the unbound oligonucleotides, the electrode was washed with 0.05 M PBS, pH 7.0. The development of DNA microarrays was provided by bounding the negatively charged phosphate backbone of the DNA probes with a positively charged from amine groups of l-cysteine. The stability of the probe through the ionic attraction achieved via applying the potential during immobilization process [31, 32].

DNA hybridization

Most of electrochemical DNA biosensor was developed based on separate processing of DNA hybridization and indicator intercalation. In the current study, the proposed DNA biosensor was preformed according to the one-step of DNA hybridization and indicator intercalation detection.

Hybridization and labeling event was performed by soaking the immobilized HPV18 DNA probe into the solution containing the complementary DNA (cDNA) and AQMS in 0.05 M PBS (pH 7.0) for 15 min at 42 Â°C. The 0.05 M PBS pH 7.0 was used to remove the nonhybridized oligonucleotides. For the interaction of HPV18 probe with noncomplementary DNA, the same process was applied. In order to the hybridization of the genomic DNA with HPV18 DNA probe on the modified SPCE, the extracted DNA from the real sample was denatured by heating in a water bath (95 Â°C) for 10 min and then placed at room temperature to obtain denatured ss-DNA. The SPCE was washed with 0.05 M PBS, pH 7.0 to remove the nonhybridized sequences. The schematic process of designing the proposed HPV 18 DNA biosensor including the immobilization and hybridization of DNA oligomers is illustrated in Fig. 2.

Fig. 2
figure 2

The proposed HPV 18 DNA biosensor design based on the immobilization of the ssDNA probe, DNA hybridization, and indicator intercalation detection

Electrochemical measurements

The Electrochemical analysis were studied by DPV method. The electrochemical measurement of ssDNA probes immobilized l-cys-AuNPs composite SPCE with AQMS immobilized in cDNA was performed in 0.05 M PBS, pH 7.0, under the potential sweep between − 0.7 and − 0.2 V with a scan rate of 50 mV s−1.

Results and discussion

Preliminary studies

In this study, the hybridization of DNA probe and target DNA was studied by UV–Vis spectroscopy and electrochemical oxidation of AQMS. One of the common methods for detection of nucleic acid is an absorbance measurement via UV–Vis spectroscopy. AQMS as an anionic redox intercalator usually used for determination of DNA hybridization [33, 34]. In this study, the mechanism of HPV 18 DNA biosensor based on DNA hybridization sensing was studied through the AQMS electrostatic adsorbtion onto the sensing element of double-stranded DNA.

For the UV–Vis study, the modified SPCE was soaked in a solution contained AQMS redox-active and cDNA. After a period of time, the UV–Vis technique was applied to the modified SPCE to investigate the surface species. As illustrated in Fig. 3, the AQMS showed a strong absorption spectra at 295 nm as well as a weakly absorption spectra at 225 nm of DNA associated with n → π* and π → π* transition state. A high absorbance signal with a slightly red-shift of spectra would be observed from the dsDNA modified electrode. This is confirmed by the action of AQMS as an intercalator in the redox signal enhancement through inserting between the double-helix structures of the dsDNA based on the help of its planar aromatic ring. Moreover, no DNA peak absorption at 225 nm, which indicated that the DNA was hybridized successfully.

Fig. 3
figure 3

UV-Vis absorbance spectra of DNA probe, cDNA and their hybridization containing 1 mM AQMS in 0.05 M PBS at pH 7.0

The oxidation and reduction products of two hydrogen ions and two electron transfers from AQMS reagent can easily carry throughout the elongated and rigid structure of double-stranded DNA which can be measure as a current charge by DPV method [35, 36] (Scheme 2).

Scheme 2
scheme 2

The proposed mechanism for electrochemical reaction of AQMS

The electrodynamic characteristics of modified SPCE with l-cysteine-AuNPs and rGO-MWCNTs were obtained by means of differential pulls voltammetric method in PBS containing AQMS. The potential difference of oxidation/reduction of AQMS (Ep) and electrical current flow (ip) were determined based on Nicholson-Shain and Randles–Sevcik equations, respectively as follows [37]:

$$E_{P} = E^{o} - \left( {\frac{RT}{3nF}} \right)\ln \left[ {\frac{{4.78\varPi 3D_{O} }}{{2D_{R} }}} \right] - \left( {\frac{RT}{3nF}} \right)\ln \left( {\frac{\alpha }{{kC_{O}^{*} }}} \right)$$
(3)
$$i_{p} = (2.69 \times 10^{5} )n^{{{3 \mathord{\left/ {\vphantom {3 2}} \right. \kern-0pt} 2}}} .A.C^{*} .D^{{{1 \mathord{\left/ {\vphantom {1 2}} \right. \kern-0pt} 2}}} .v^{{{1 \mathord{\left/ {\vphantom {1 2}} \right. \kern-0pt} 2}}}$$
(4)

Where, k is the electron transfer rate constant, C* is the bulk concentration of reductive/oxidative species, Ê‹ is the scan rate, DR and DO are the diffusion coefficient of the reduced and oxidized species, respectively.

As it is obvious in several reports, a sharp current peak at ≈ − 0.43 V resulted from measuring the DPV response of proposed HPV 18 biosensor was noticed.

Figure 4 shows the DPV voltammograms for the AQMS electrochemical oxidation of bare SPCE (a), modified electrode with l-cysteine-AuNPs (b), modified electrode with l-cysteine-AuNPs and rGO (c), modified electrode with l-cysteine-AuNPs and rGO-MWCNTs (d) in 0.05 M PBS (pH 6.5) with a scan rate of 50 mV s−1. As seen in Fig. 4, it was observed that the bare SPCE showed the lowest current peak which represents a poor electron transfer [19]. In contrast, modifying the SPCE with l-cysteine-AuNPs showed a slight increase in the current signal rate of about 2 μA due to AQMS oxidation at SPCE-solution interface via offering catalytic properties and excellent conductivity of AuNPs [38]. The highest current peak is related to SPCE modified with l-cysteine-AuNPs and rGO-MWCNTs. This is confirmed that rGO-MWCNT nanocomposite has a remarkable effect on the sensitivity of the modified electrode due to its excellent conductivity and high surface-to-volume ratio via promoting and possessing the electron transfer and rapid electrode kinetics, respectively [39, 40]. In addition, the obtained results proved that the AuNPs as excellent electric conductive materials provide a suitable surface for immobilization of thiolated molecules through chemisorptions. The electric conductive molecules such as MWCNTs and rGO containing both thiol (supplied by l-cysteine) and carboxyl groups can be used as linkers for functionalized AuNPs. The solubility of rGO in aqueous solutions provides its wide application in biological samples.

Fig. 4
figure 4

Differential pulse voltammograms for the AQMS electrochemical oxidation of bare SPCE (a), a modified electrode with l-cysteine-AuNPs (b), a modified electrode with l-cysteine-AuNPs and rGO (c) and a modified electrode with l-cysteine-AuNPs and rGO-MWCNTs (d). The DPV analysis was conducted in 0.05 M PBS (PH 6.5) with a scan rate of 50 mV s−1 versus Ag/AgCl reference electrode

Optimization and characterization of HPV 18 DNA biosensor

Effect of DNA probe

In this study, the electrochemical behavior of the probe concentration on the modified SPCE through AQMS electrochemical oxidation was investigated. As shown in Fig. 5a, the effect of probe concentration was obtained by using DPV technique in 0.05 M PBS (pH = 7). The results demonstrated that the concentrations of 0.01 Î¼M and 0.1 Î¼M indicate the lowest signals due to the low amount of DNA probe immobilized on the surface of modified SPCE. The concentration of 5 Î¼M showed the highest current peak which resulted in the most hybridization of target cDNA. Also, no increase in signal strength was observed for a concentration of 10 Î¼M or higher. It can be explained by the lower availability of the intercalator AQMS and massive probe accumulation on the surface of SPCE that causes probe overloading. As a result, the concentration of 5 Î¼M was determined as the optimal concentration for DNA probes [41, 42].

Fig. 5
figure 5

Histogram of (a) the HPV18 probe concentration effect on the AQMS electrochemical oxidation during immobilization process in 0.05 M PBS pH 7. The electrochemical analysis was studied by DPV method in the following conditions: initial potential − 0.7 V, end potential − 0.2 V, modulation amplitude 50 mV and scan rate 50 mVs−1. All results were performed with triplicate measurement of each HPV18 probe concentration, and (b) DPV oxidation current signal of HPV-ssDNA immobilized at the surface of SPCE during hybridization with DNA Target, DNA one mismatch, DNA three mismatch and non-complimentary DNA in 0.05 M PBS at pH 7 with the scan rate of 50 mVs−1

Effect of hybridization for synthetic HPV 18 DNA target

The DNA target was detected through the DPV responses of the modified biosensor to the electrochemical oxidation of AQMS. For determining the selectivity of biosensor response, the DNA probe hybridization with nil DNA target, DNA target, DNA one mismatch, DNA three mismatch and DNA non-complementary was investigated. In Fig. 5b, the highest current peak is related to the hybridization of the probe-modified electrode and cDNA but unlike that, the lower current is due to the use of non-complementary DNA. Also, the use of one mismatch DNA and three mismatch DNA results in lower signal flow, due to the lack of complete hybridization. This can be explained due to more AQMS accumulation on the ds-DNA caused by the affinity of AQMS to the double-helix structure of the dsDNA.

Calibration curve

To identify the conditions of optimized HPV-18 DNA biosensor, the oxidation of AQMS was measured during hybridization of DNA probes with different concentrations of synthetic DNA targets (0.01fM to 0.01 nM) by the DPV method. Figure 6 shows that with increasing concentration of DNA target, an increasing trend in the current signal was observed. It is thus believed that by increasing the concentration of DNA target and subsequently increasing their hybridization with the DNA probe, AQMS reduction increases peak current. A calibration curve of the current signals (ΔI) various the concentrations of targets DNA in from the DPV analysis was plotted. As it is obvious from Fig. 6, the calibration curve exhibited a linear dynamic range in a concentration range of 0.01 fM to 0.01 nM with a correlation coefficient of 0.994. Moreover, the DPV peak potential shows a slightly negative shifts by about 10 mV with the increase of the DNA target concentration in the studied concentration range. Presumably, this shift is due to the Donnan equilibria in the nanocomposite film [43, 44].

Fig. 6
figure 6

The differential pulse voltammograms and calibration plots for the AQMS electrochemical reduction after hybridization with different concentration of DNA target (a: 10−8, b: 10−7, c: 10−6, d: 10−5, e: 10−4, f: 10−3 and g: 10−2 nM) in 0.05 M PBS pH 7. The DPV method was applied under the following conditions: initial potential − 0.7 V, end potential − 0.2 V, modulation amplitude 50 mV and scan rate 50 mVs−1

The derived limit of quantitation (LOQ) (at 10SD/α) and limit of detection (LOD) (at 3SD/α) [16, 45] of the proposed HPV-18 DNA biosensor were calculated to be 0.16 and 0.05 fM, respectively. In these equations, SD is defined as the standard deviation of the intercept and α is defined as the slope of the linear regression. The stability and repeatability of the proposed biosensor were also examined. The relative standard deviation of the five repetitive measurements was 1.30%, which is less than 20%, demonstrating the good reusability and stability of the proposed biosensor.

FTIR-ATR analysis

The FTIR-ATR spectra of functional groups of modified SPCE with (a) l-cysteine-AuNPs and rGO-MWCNTs (b) ssDNA probe and (c) dsDNA exhibits a peak series in the region of 4000–500 cm−1 (Fig. 7). The FTIR-ATR spectra of rGO-MWCNT/L-Cys-AuNPs composite shows important peaks at 3370 cm−1, 2974 cm−1, 1452 cm−1 and 750-665 cm−1 for (O–H stretch), (S–H stretch), (C = O stretch) and (C-S; Au–S stretches), respectively (a). The formation of self-assembled monolayer modified SPCE (l-Cys/Au SAMs) was obtained through combining the l-Cys onto AuNPs by taking advantage of strong Au–S interaction. The immobilization of ssDNA probe onto a modified SPCE with rGO-MWCNT/l-Cys-AuNPs lead to a decrease in peak intensity related to Au–S, C–S and O–H stretching (b). A FT-IR spectrum of ssDNA HPV-18 after hybridization with genomic DNA exhibits similar peaks of different bases at the same position with an increase in peak intensity (c). The N–H bending and stretching of purine and pyrimidine ring as DNA bases, is corresponded to the presented peaks at 3355 cm−1 [35]. The presented results show the successful process of immobilization and hybridization of the ssDNA probe with genomic DNA.

Fig. 7
figure 7

FT-IR spectra of (a) modified SPCE with l-cysteine-AuNPs and rGO-MWCNTs (b) ssDNA HPV-18 probe immobilized SPCE and (c) ssDNA hybridized SPCE

FESEM analysis

To confirm the rGO-MWCNT/AuNPs composites stabilization on the surface of SPCE, the field-emission scanning electron microscopy (FESEM) examination was used [12]. Figure 8a, b, and c are FESEM nano-graphs of the bare SPCE, rGO-MWCNT/AuNPs composite modified SPCE with and without l-cysteine, respectively. As it is shown in Fig. 8a, the graphitic carbon powder is responsible for the small particles dispersed throughout the bare SPCE which assigned the sheet-like structures. After modifying the SPCE with rGO-MWCNT composite, the presence of MWCNTs which bridge and covered the rGO sheets and small particles of graphitic powder were observed. The presence of AuNps which improves the electrocatalytic activity by enhancing the conductivity for fast electron transfer, is seen that distributed on the rGO-MWCNT network in Fig. 8b, c. Moreover, as shown in Fig. 8b, when AuNPs are deposited on the SPCE surface modified with rGO-MWCNT, a partial accumulation of them is observed which is partly due to the strong Van der Waals force between the nanomaterial. In the following, after electrodeposition of l-cysteine on the modified SPCE, the accumulation of AuNPs decreases and the particle size increases (Fig. 8c) [24].

Fig. 8
figure 8

FESEM image of (a) bare SPCE (scale 1000 nm), (b) modified SPCE with rGO-MWCNTs/AuNPs (scale 200 nm) and (c) modified SPCE with rGO-MWCNTs/AuNPs-l-cysteine (scale 500 nm)

EIS analysis

In this section, the EIS was applied as an effective technique for investigation of the proposed electrode behavior after each assembly steps. The EIS recorded in 1 × 10−3 M [Fe(CN)6]−3/[Fe(CN)6]−4 (1:1) containing 0.1 M KCl solution in the frequency range of 0.1–100 kHz followed by applying a signal amplitude of the open circuit potential. As it is obvious in Fig. 9, the Nyquist plots of SPCE modified with (a) rGO-MWCNTs, (b) rGO-MWCNTs/AuNPs, (c) rGO-MWCNTs/AuNPs-l-cysteine-ssDNA probe and (d) rGO-MWCNTs/AuNPs-l-cysteine-ssDNA probe with DNA target. The SPCE modified with rGO-MWCNTs showed a resistance (14.8 kΩ), which indicated that rGO-MWCNTs was an electric conducting material and accelerated the electron transfer. The Rct decreased evidently (797 Î©) because AuNPs loaded on the surface of SPCE/rGO-MWCNTs modified electrode formed an excellent carrier and accelerated the further electron transfer. The loading of the ssDNA probe on the modified SPCE led to the enhancement in Rct up to 800 Î©. Significant enhancement Rct (37.2 kΩ) by hybridization of ssDNA probe with ssDNA target formed a barrier and further prevented the redox probe to the electrode surface (Fig. 9).

Fig. 9
figure 9

Nyquist plots and equivalent circuit models of the impedance spectra recorded for SPCE modified with (a) rGO-MWCNTs, (b) rGO-MWCNTs/AuNPs, (c) rGO-MWCNTs/AuNPs-l-cysteine-ssDNA probe and (d) rGO-MWCNTs/AuNPs-l-cysteine-ssDNA probe with DNA target

Performance of proposed biosensor in the analysis of HPV in real sample

The performance of proposed HPV-18 DNA biosensor was investigated via applying it in the extracted DNA samples from human patient. For this purpose, the PCR was performed and the amplification products were run in the gel electrophoresis [46, 47]. As shown in Fig. 10, the results obtained from the positive DNA HPV-18 patient samples extracted from the real samples were consistent with the results of synthesized DNA. An increment in signal strength was observed with increasing target DNA concentration. In addition, the DPV signal of hybridization of the probe modified-SPCE with negative HPV-18 patient sample decreased in compare to the positive forms, indicating that there is a poor hybridization signal in the negative patient sample.

Fig. 10
figure 10

a DNA extracted from clinical sample and presence of DNA of HPV type 18 virus on gel electrophoresis; b the differential pulse voltammograms for the AQMS electrochemical reduction after hybridization with DNA extract from the real sample. The electrochemical analysis was studied by DPV method in the following conditions: initial potential − 0.7 V, end potential − 0.1 V, modulation amplitude 50 mV and scan rate 50 mVs−1

These data demonstrated that the proposed HPV-18 DNA electrochemical biosensor able to detect the HPV type 18 in PCR real samples.

Standard synthesized target ssDNA HPV18 has been spiked into the negative cervical samples within the calibration concentration range of the biosensor for recovery of target ssDNA HPV18 in the cervical samples, and 0.05 M PBS (pH 7.0) was used to stabilize the pH level of the samples. From the results tabulated in Table 3, satisfactory recoveries between 101.8 and 128% were obtained through applying the proposed DNA biosensor for accurate and reliable target DNA HPV18 determination in the cervical samples.

Table 3 Recovery of synthesized target DNA HPV18 in the real samples

Comparison of proposed HPV-18 DNA biosensor with other studies

The analytical performance of current biosensor was compared with previous published HPV-18 DNA electrochemical biosensors and the results were listed in Table 4. The proposed DNA biosensor based on a SPCE modified with a nanocomposite of rGO-MWCNT/AuNPs showed an improvement in the linearity response range in compare to previous works. It is also compared favorably with other reported HPV-18 DNA biosensors in terms of the lower detection limit. Therefore, rGO-MWCNT/AuNPs modified SPE was confirmed to be an efficient platform for the electrochemical sensor. It might be mainly ascribed to the large specific surface area of the rGO-MWCNT/AuNPs nanocomposite that improved the immobilization amount of ssDNA probes and the fine electronic conduction ability of AQMS that improved the detection signals [13]. Moreover, the formation of self-assembled monolayer modified SPCE (l-Cys/Au SAMs) enhanced the electrochemical responses towards DNA targets and also provides the less prone of DNA probe to nonspecific adsorption onto the SPE surface [48]. A detailed comparison of electrochemical HPV-18 DNA biosensor reported in the literature is shown in Table 4.

Table 4 Comparison of HPV-18 DNA electrochemical biosensor with other previous reported electrochemical HPV-18 DNA biosensors

Conclusion

In this study, we designed an electrochemical biosensor to detect HPV-18, one of the most hazardous types of HPVs. The performance of this biosensor was evaluated using synthetic DNA and DNA extracted from HPV 18 patients. The proposed biosensor demonstrated successful performance in identifying the HPV extracted from real samples. The biosensor was also able to detect very small quantities and a large range of analytes due to the use of rGO/MWCNT, Au nanoparticle, l-cysteine to modify the surface of the electrode. The specific response of the biosensor was examined by a test and showed a very good result. Therefore, the designed biosensor can be used as a successful tool for rapid, accurate, easy, and early detection of HPV 18 and alternative traditional and current methods.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

HPV:

human papillomavirus

MWCNTs:

multiwalled carbon nanotubes

SPCE:

screen-printed carbon electrode

AuNP:

Au nanoparticle

ssDNA:

single strand DNA

Cys:

l-cysteine

DPV:

differential pulse voltammetry

AQMS:

anthraquninone-2-sulfonic acid monohydrate sodium salt

PCR:

polymerase chain reaction

EIS:

electrochemical impedance spectroscopy

LCD:

liquid–crystal displayer

CV:

cyclic voltammetry

LOQ:

limit of quantitation

LOD:

limit of detection

FESEM:

field-emission scanning electron microscopy

SWV:

square wave voltammetry

HRP:

horseradish peroxide

References

  1. Rasouli E, Shahnavaz Z, Basirun WJ, Rezayi M, Avan A, Ghayour-Mobarhan M, Khandanlou R, Johan MR. Advancements in electrochemical DNA sensor for detection of human papilloma virus? A review. Anal Biochem. 2018. https://doi.org/10.1016/j.ab.2018.07.002.

    Article  PubMed  Google Scholar 

  2. Yim E-K, Park J-S. The role of HPV E6 and E7 oncoproteins in HPV-associated cervical carcinogenesis. Cancer Res Treat. 2005;37(6):319.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Shah SS, Senapati S, Klacsmann F, Miller DL, Johnson JJ, Chang H-C, Stack MS. Current technologies and recent developments for screening of HPV-associated cervical and oropharyngeal cancers. Cancers. 2016;8(9):85.

    Article  PubMed Central  CAS  Google Scholar 

  4. Didelot MN, Boulle N, Damay A, Costes V, Segondy M. Comparison of the PapilloCheck® assay with the digene HC2 HPV DNA assay for the detection of 13 high-risk human papillomaviruses in cervical and anal scrapes. J Med Virol. 2011;83(8):1377–82.

    Article  PubMed  CAS  Google Scholar 

  5. Abreu AL, Souza RP, Gimenes F, Consolaro ME. A review of methods for detect human Papillomavirus infection. Virol J. 2012;9(1):262.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  6. Bhatla N, Moda N. The clinical utility of HPV DNA testing in cervical cancer screening strategies. Indian J Med Res. 2009;130(3):261.

    PubMed  CAS  Google Scholar 

  7. Cho EJ, Do JH, Kim YS, Bae S, Ahn WS. Evaluation of a liquid bead array system for high-risk human papillomavirus detection and genotyping in comparison with Hybrid Capture II, DNA chip and sequencing methods. J Med Microbiol. 2011;60(2):162–71.

    Article  PubMed  CAS  Google Scholar 

  8. Tosar J, Branas G, Laíz J. Electrochemical DNA hybridization sensors applied to real and complex biological samples. Biosensors Bioelectron. 2010;26(4):1205–17.

    Article  CAS  Google Scholar 

  9. Arduini F, Micheli L, Moscone D, Palleschi G, Piermarini S, Ricci F, Volpe G. Electrochemical biosensors based on nanomodified screen-printed electrodes: recent applications in clinical analysis. TrAC Trends Anal Chem. 2016;79:114–26.

    Article  CAS  Google Scholar 

  10. Privett BJ, Shin JH, Schoenfisch MH. Electrochemical sensors. Anal Chem. 2010;82(12):4723–41.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  11. Ronkainen NJ, Halsall HB, Heineman WR. Electrochemical biosensors. Chem Soc Rev. 2010;39(5):1747–63.

    Article  PubMed  CAS  Google Scholar 

  12. Zhang Y, Ji Y, Wang Z, Liu S, Zhang T. Electrodeposition synthesis of reduced graphene oxide–carbon nanotube hybrids on indium tin oxide electrode for simultaneous electrochemical detection of ascorbic acid, dopamine and uric acid. RSC Adv. 2015;5(129):106307–14.

    Article  CAS  Google Scholar 

  13. Zhu Z. An overview of carbon nanotubes and graphene for biosensing applications. Nano-Micro Lett. 2017;9(3):25.

    Article  CAS  Google Scholar 

  14. Tîlmaciu C-M, Morris MC. Carbon nanotube biosensors. Front Chem. 2015;3:59.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  15. Li Y, Schluesener HJ, Xu S. Gold nanoparticle-based biosensors. Gold Bull. 2010;43(1):29–41.

    Article  Google Scholar 

  16. Nascimento GA, Souza EV, Campos-Ferreira DS, Arruda MS, Castelletti CH, Wanderley MS, Ekert MH, Bruneska D, Lima-Filho JL. Electrochemical DNA biosensor for bovine papillomavirus detection using polymeric film on screen-printed electrode. Biosensors Bioelectron. 2012;38(1):61–6.

    Article  CAS  Google Scholar 

  17. Umuhumuza LC, Sun X. Rapid detection of pork meat freshness by using l-cysteine-modified gold electrode. Eur Food Res Technol. 2011;232(3):425–31.

    Article  CAS  Google Scholar 

  18. Ferreon ACM, Moran CR, Gambin Y, Deniz AA. Single-molecule fluorescence studies of intrinsically disordered proteins. Methods Enzymol. 2010;472:179–204.

    Article  PubMed  CAS  Google Scholar 

  19. Campos-Ferreira DS, Nascimento GA, Souza EV, Souto-Maior MA, Arruda MS, Zanforlin DM, Ekert MH, Bruneska D, Lima-Filho JL. Electrochemical DNA biosensor for human papillomavirus 16 detection in real samples. Anal Chim Acta. 2013;804:258–63.

    Article  PubMed  CAS  Google Scholar 

  20. Fischer S, Bettstetter M, Becher A, Lessel M, Bank C, Krams M, Becker I, Hartmann A, Jagla W, Gaumann A. Shift in prevalence of HPV types in cervical cytology specimens in the era of HPV vaccination. Oncol Lett. 2016;12(1):601–10.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  21. Hasanzadeh M, Rejali M, Mehramiz M, Akbari M, Mousavi Seresht L, Yazdandoost Y, Farokhi S, Mahdian Z, Maleki F, Emamdadi-Aliabad Z. The interaction of high and low-risk human papillomavirus genotypes increases the risk of developing genital warts: a population-based cohort study. J Cell Biochem. 2019;120:12870–4.

    Article  PubMed  CAS  Google Scholar 

  22. Roodan VA, Hosseinpour N, Zaman M. Electrochemical properties of electrodeposited reduced graphene oxide and carbon nanotubes. J Nanosci Nanotechnol. 2017;17(4):2493–502.

    Article  PubMed  CAS  Google Scholar 

  23. Jia F, Duan N, Wu S, Dai R, Wang Z, Li X. Impedimetric salmonella aptasensor using a glassy carbon electrode modified with an electrodeposited composite consisting of reduced graphene oxide and carbon nanotubes. Microchim Acta. 2016;183(1):337–44.

    Article  CAS  Google Scholar 

  24. Zhang J, Tan W, Tao Y, Deng L, Qin Y, Kong Y. A novel electrochemical chiral interface based on sandwich-structured molecularly imprinted SiO2/AuNPs/SiO2 for enantioselective recognition of cysteine isomers. Electrochem Commun. 2018;86:57–62.

    Article  CAS  Google Scholar 

  25. He J, Xing S, Qiu R, Song Z, Zhang S. l-Cysteine/glycine composite film modified glassy carbon electrode as an enhanced sensing platform for catechol determination. Anal Methods. 2014;6(4):1210–8.

    Article  CAS  Google Scholar 

  26. Niu H, Yuan R, Chai Y, Mao L, Yuan Y, Zhuo Y, Yuan S, Yang X. Electrochemiluminescence of peroxydisulfate enhanced by l-cysteine film for sensitive immunoassay. Biosensors Bioelectron. 2011;26(7):3175–80.

    Article  CAS  Google Scholar 

  27. Ajayi OA, Guitierrez DH, Peaslee D, Cheng A, Gao T, Wong CW, Chen B. Electrophoretically deposited graphene oxide and carbon nanotube composite for electrochemical capacitors. Nanotechnology. 2015;26(41):415203.

    Article  PubMed  CAS  Google Scholar 

  28. Bittolo Bon S, Valentini L, Kenny J, Peponi L, Verdejo R, Lopez-Manchado M. Electrodeposition of transparent and conducting graphene/carbon nanotube thin films. Physica Status Solidi (a). 2010;207(11):2461–6.

    Article  CAS  Google Scholar 

  29. Ahmad H, Aidit S, Ooi S, Rezayi M, Tiu Z. Passively Q-switched and mode-locked erbium doped fiber laser based on N-doped graphene saturable absorber. Laser Phys. 2017;27(10):105302.

    Article  CAS  Google Scholar 

  30. Ahmad H, Safaei R, Rezayi M, Amiri I. Novel D-shaped fiber fabrication method for saturable absorber application in the generation of ultra-short pulses. Laser Phys Lett. 2017;14(8):085001.

    Article  Google Scholar 

  31. Sethi M, Pacardo DB, Knecht MR. Biological surface effects of metallic nanomaterials for applications in assembly and catalysis. Langmuir. 2010;26(19):15121–34.

    Article  PubMed  CAS  Google Scholar 

  32. Rashid JIA, Yusof NA. The strategies of DNA immobilization and hybridization detection mechanism in the construction of electrochemical DNA sensor: a review. Sensing Bio-Sensing Res. 2017;16:19–31.

    Article  Google Scholar 

  33. Ulianas A, Heng LY, Hanifah SA, Ling TL. An electrochemical DNA microbiosensor based on succinimide-modified acrylic microspheres. Sensors. 2012;12(5):5445–60.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  34. Pan D, Zuo X, Wan Y, Wang L, Zhang J, Song S, Fan C. Electrochemical interrogation of interactions between surface-confined DNA and methylene blue. Sensors. 2007;7(11):2671–80.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  35. Yuan G, Zhang G, Chen J, Fu L, Xu L, Yang F. The electrochemical activities of anthraquinone monosulfonate adsorbed on the basal plane of reduced graphene oxide by π–π stacking interaction. J Solid State Electrochem. 2013;17(10):2711–9.

    Article  CAS  Google Scholar 

  36. Feng F, Uno B, Goto M, Zhang Z, An D. Anthraquinone-2-sulfonyl chloride: a new versatile derivatization reagent—synthesis mechanism and application for analysis of amines. Talanta. 2002;57(3):481–90.

    Article  PubMed  CAS  Google Scholar 

  37. Kaczmarska K, Brycht M, Leniart A, Skrzypek S. Differential pulse voltammetric determination of an immunosuppressive drug teriflunomide on an edge plane pyrolytic graphite electrode. RSC Adv. 2017;7(42):26028–36.

    Article  CAS  Google Scholar 

  38. Muhammad A, Hajian R, Yusof NA, Shams N, Abdullah J, Woi PM, Garmestani H. A screen printed carbon electrode modified with carbon nanotubes and gold nanoparticles as a sensitive electrochemical sensor for determination of thiamphenicol residue in milk. RSC Adv. 2018;8(5):2714–22.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. An KH, Kim WS, Park YS, Moon JM, Bae DJ, Lim SC, Lee YS, Lee YH. Electrochemical properties of high-power supercapacitors using single-walled carbon nanotube electrodes. Adv Funct Mater. 2001;11(5):387–92.

    Article  CAS  Google Scholar 

  40. Wang Z, Dai Z. Carbon nanomaterial-based electrochemical biosensors: an overview. Nanoscale. 2015;7(15):6420–31.

    Article  PubMed  CAS  Google Scholar 

  41. Pournaghi-Azar MH, Hejazi MS, Alipour E. Developing an electrochemical deoxyribonucleic acid (DNA) biosensor on the basis of human interleukine-2 gene using an electroactive label. Anal Chim Acta. 2006;570(2):144–50.

    Article  PubMed  CAS  Google Scholar 

  42. Doneux T, De Rache A, Triffaux E, Meunier A, Steichen M, Buess-Herman C. Optimization of the probe coverage in dna biosensors by a one-step coadsorption procedure. ChemElectroChem. 2014;1(1):147–57.

    Article  CAS  Google Scholar 

  43. Ferancová A, Korgová E, Buzinkaiová T, Kutner W, Štěpánek I, Labuda J. Electrochemical sensors using screen-printed carbon electrode assemblies modified with the β-cyclodextrin or carboxymethylated β-cyclodextrin polymer films for determination of tricyclic antidepressive drugs. Anal Chim Acta. 2001;447(1–2):47–54.

    Article  Google Scholar 

  44. Kutner W, Doblhofer K. Simultaneous cyclic voltammetry and electrochemical quartz-crystal microbalance study at polymer film-modified electrodes of molecular inclusion of ferrocene by β-cyclodextrin polymer and carboxymethylated β-cyclodextrin polymer as well as ferrocenecarboxylic acid by β-cyclodextrin polymer. J Electroanal Chem. 1992;326(1–2):139–60.

    Article  CAS  Google Scholar 

  45. Gumustas M, Ozkan SA. The role of and the place of method validation in drug analysis using electroanalytical techniques. Open Anal Chem J. 2011;5(1):1–21.

    Article  CAS  Google Scholar 

  46. Gao H, Qi X, Chen Y, Sun W. Electrochemical deoxyribonucleic acid biosensor based on the self-assembly film with nanogold decorated on ionic liquid modified carbon paste electrode. Anal Chim Acta. 2011;704(1–2):133–8.

    Article  PubMed  CAS  Google Scholar 

  47. Wu G, Yang N, Zhang T, Wang Z, Lu X, Kang J. Fabrication and application of a new DNA biosensor based on on-substrate PCR and electrochemistry. Sensors Actuators B Chem. 2011;160(1):598–603.

    Article  CAS  Google Scholar 

  48. Huang H, Bai W, Dong C, Guo R, Liu Z. An ultrasensitive electrochemical DNA biosensor based on graphene/Au nanorod/polythionine for human papillomavirus DNA detection. Biosensors Bioelectron. 2015;68:442–6.

    Article  CAS  Google Scholar 

  49. Civit L, Fragoso A, Hölters S, Dürst M, O’Sullivan CK. Electrochemical genosensor array for the simultaneous detection of multiple high-risk human papillomavirus sequences in clinical samples. Anal Chim Acta. 2012;715:93–8.

    Article  PubMed  CAS  Google Scholar 

  50. Kowalczyk A, Nowicka AM. Application of mercury–mediated thymine–base pairs for successful voltammetric detection of HPV 18. Sensors Actuators B Chem. 2016;237:810–6.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors thank financial supports from University of Medical Sciences, Mashhad, Iran and the National Institutes for Medical Research Development (NIMAD), Tehran, Iran for Elite Researcher Grant Committee under award number 963622.

Funding

This work was supported by University of Medical Sciences, Mashhad, Iran.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the conception and design of the study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Majid Rezayi.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mahmoodi, P., Rezayi, M., Rasouli, E. et al. Early-stage cervical cancer diagnosis based on an ultra-sensitive electrochemical DNA nanobiosensor for HPV-18 detection in real samples. J Nanobiotechnol 18, 11 (2020). https://doi.org/10.1186/s12951-020-0577-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12951-020-0577-9

Keywords