Abstract
Background
Multiple cases of dissociation of the head from the neck termed as gross trunnion failure (GTF) in total hip replacement have been described. Very little quantitative data is available for patient and implant factors associated and predisposing to this complication.
Study purpose
To systematically review and analyze all studies which have gross trunnion failure in case of metal on polyethylene (MoP) total hip replacement.
Methods
PubMed database was searched. We also performed a secondary search by pearling bibliography of all full text articles obtained. Predefined inclusion and exclusion criteria were used for abstract screening by two independent observers. A total 46 cases met our inclusion criteria. These were included in the final analysis and data was pooled.
Results
Till date, 46 cases of GTF in MoP THR have been reported. The mean age at time of revision was 70.13 years (range 50 to 89 years). The mean time to revision surgery was 8.24 years (range 4.7 to 14 years). 91.4% cases were male. BMI was ≥ 25 in 38/41 cases and ≥ 30 in 21/41 cases. Pain (95.5%) and difficulty or inability to walk (97.7%) were the most common symptoms. A total of 19/44 cases described varying sounds like click, clunk, and pop before dislocation. Accolade TMZF/TMZF plus was the most common stem used in 34/46 cases. The stem neck angle was 127 degrees in 97% cases (32/33 cases). A positive neck offset of 4 mm or more was used in 91.1% cases (40/44 cases). The head size ≥ 36 mm in approximately 90% cases. The head material was cobalt chromium in all 45 cases, where data was available. Among the intra-operative findings, the most common findings were metallosis (41/44), black or brow coloured synovial fluid (21/44), pseudotumour (19/44), synovial hypertrophy (18/44), damage to the abductor musculature (17/44), proximal femoral osteolysis (5/44 cases), and heterotrophic ossification (2/44 cases). Histopathological analysis was available in 11 cases and was suggestive of fibrous tissue with chronic cellular infiltrate in all cases. The serum cobalt and chromium concentrations were raised above normal in 86.4% (19/22) and 21.7% cases (5/23) respectively.
Conclusions
Gross trunnion failure may not be as uncommon as was previously thought. A number potential associations and predisposing factors of this complication have been highlighted in this review. But due to small sample size and weak level of evidence, further studies are needed in this field.
Similar content being viewed by others
References
Lanting BA, Teeter MG, Howard JL, MacDonald SJ, Van Citters DW (2017) Metal-on-metal compared with metal-on-polyethylene: the effect on trunnion corrosion in total hip arthroplasty. J Arthroplast 32(8):2574–2579. https://doi.org/10.1016/j.arth.2017.03.012
Lucas LC, Buchanan RA, Lemons JE (1981) Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res 15(5):731–747. https://doi.org/10.1002/jbm.820150509
Urish KL, Giori NJ, Lemons JE, Mihalko WM, Hallab N (2019) Trunnion corrosion in total hip arthroplasty—basic concepts. Orthop Clin N Am 50(3):281–288
Pennock AT, Schmidt AH, Bourgeault CA (2002) Morse-type tapers: factors that may influence taper strength during total hip arthroplasty. J Arthroplast 17(6):773–778. https://doi.org/10.1054/arth.2002.33565
Gill IPS, Webb J, Sloan K, Beaver RJ (2012) Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg (Br) 94(7):895–900. https://doi.org/10.1302/0301-620X.94B7.29122
Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JP et al (2009) The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. PLoS Med 6(7):e1000100. https://doi.org/10.1371/journal.pmed.1000100
Ko LM, Chen AF, Deirmengian GK, Hozack WJ, Sharkey PF (2016) Catastrophic femoral head-stem trunnion dissociation secondary to corrosion. J Bone Joint Surg Am 98(16):1400–1404. https://doi.org/10.2106/JBJS.15.00914
Raju S, Chinnakkannu K, Puttaswamy MK, Phillips MJ (2017) Trunnion corrosion in metal-on-polyethylene total hip arthroplasty: a case series. J Am Acad Orthop Surg 25(2):133–139. https://doi.org/10.5435/JAAOS-D-16-00352
Banerjee S, Cherian JJ, Bono JV, Kurtz SM, Geesink R, Meneghini RM et al (2015) Gross trunnion failure after primary total hip arthroplasty. J Arthroplast 30(4):641–648. https://doi.org/10.1016/j.arth.2014.11.023
Talmo CT, Sharp KG, Malinowska M, Bono JV, Ward DM, Lareau J (2014) Spontaneous modular femoral head dissociation complicating total hip arthroplasty. Orthopedics 37(6):e592–e595. https://doi.org/10.3928/01477447-20140528-62
Parker SJ, Khan W, Mellor S (2015) Late nontraumatic dissociation of the femoral head and trunnion in a total hip arthroplasty. Case Rep Orthop 2015. https://doi.org/10.1155/2015/738671
Urish KL, Hamlin BR, Plakseychuk AY, Levison TJ, Higgs GB, Kurtz SM, DiGioia AM (2017) Trunnion failure of the recalled low friction ion treatment cobalt chromium alloy femoral head. J Arthroplast 32(9):2857–2863. https://doi.org/10.1016/j.arth.2017.03.075
Runner RP, Bellamy JL, Roberson JR (2016) Gross trunnion failure of a cobalt-chromium femoral head on a titanium stem at midterm follow-up: a report of 3 cases. JBJS Case Connect 6(4):e96. https://doi.org/10.2106/JBJS.CC.16.00054
Shamshoon S, Thornley P, de Beer J (2018) Profound trunnion wear resulting in femoral head-neck dissociation in total hip arthroplasty. Case Rep Orthop 2018. https://doi.org/10.1155/2018/1534572
Swann RP, Webb JE, Cass JR, Van Citters DW, Lewallen DG (2015) Catastrophic head-neck dissociation of a modular cementless femoral component. JBJS Case Connect 5(3):e71. https://doi.org/10.2106/JBJS.CC.N.00179
Arvinte D, Kiran M, Sood M (2015) Femoral head-trunnion dissociation secondary to corrosion of a modular total hip replacement: a case report. JBJS Case Connect 5(2):e35. https://doi.org/10.2106/JBJS.CC.N.00146
Spanyer J, Hines J, Beaumont CM, Yerasimides J (2016) Catastrophic femoral neck failure after THA with the Accolade® I stem in three patients. Clin Orthop Relat Res 474(5):1333–1338. https://doi.org/10.1007/s11999-015-4438-z
Bolarinwa SA, Martino JM, Moskal JT, Wolfe MW, Shuler TE (2019) Gross trunnion failure after a metal-on-polyethylene total hip arthroplasty leading to dissociation at the femoral head-trunnion interface. Arthroplast Today 5(1):5–10. https://doi.org/10.1016/j.artd.2019.02.001
Wood TJ, Alzahrani M, Langohr D, Teeter MG, Howard JL, Schemitsch EH, Lanting BA (2019) Catastrophic femoral head trunnion dissociation: a case series with surface wear analysis. Hip Int:1120700019860774. https://doi.org/10.1177/1120700019860774
López-López JA, Humphriss RL, Beswick AD, Thom HH, Hunt LP, Burston A, Blom AW (2017) Choice of implant combinations in total hip replacement: systematic review and network meta-analysis. BMJ 359:j4651. https://doi.org/10.1136/bmj.j4651
Jennings JM, Dennis DA, Yang CC (2016) Corrosion of the head-neck junction after total hip arthroplasty. J Am Acad Orthop Surg 24(6):349–356. https://doi.org/10.5435/JAAOS-D-15-00111
Porter DA, Urban RM, Jacobs JJ, Gilbert JL, Rodriguez JA, Cooper HJ (2014) Modern trunnions are more flexible: a mechanical analysis of THA taper designs. Clin Orthop Relat Res 472(12):3963–3970. https://doi.org/10.1007/s11999-014-3965-3
Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S (2002) A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res 401:149–161
Langton DJ, Sidaginamale R, Lord JK, Nargol AVF, Joyce TJ (2012) Taper junction failure in large-diameter metal-on-metal bearings. Bone Joint Res 1(4):56–63. https://doi.org/10.1302/2046-3758.14.2000047
Lavernia CJ, Iacobelli DA, Villa JM, Jones K, Gonzalez JL, Jones WK (2015) Trunnion–head stresses in tha: are big heads trouble? J Arthroplast 30(6):1085–1088. https://doi.org/10.1016/j.arth.2015.01.021
Bolland BJRF, Culliford DJ, Langton DJ, Millington JPS, Arden NK, Latham JM (2011) High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg (Br) 93(5):608–615. https://doi.org/10.1302/0301-620X.93B5.26309
Matthies AK, Racasan R, Bills P, Blunt L, Cro S, Panagiotidou A, Hart AJ (2013) Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res 31(11):1677–1685. https://doi.org/10.1002/jor.22431
Toms AP, Marshall TJ, Cahir J, Darrah C, Nolan J, Donell ST et al (2008) MRI of early symptomatic metal-on-metal total hip arthroplasty: a retrospective review of radiological findings in 20 hips. Clin Radiol 63(1):49–58. https://doi.org/10.1016/j.crad.2007.07.012
Haddad FS, Thakrar RR, Hart AJ, Skinner JA, Nargol AVF, Nolan JF et al (2011) Metal-on-metal bearings: the evidence so far. J Bone Joint Surg (Br) 93(5):572–579. https://doi.org/10.1302/0301-620X.93B4.26429
Mao X, Tay GH, Godbolt DB, Crawford RW (2012) Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty. J Arthroplast 27(3):493–e13. https://doi.org/10.1016/j.arth.2011.07.015
Odak S, Ivory J (2013) Management of abductor mechanism deficiency following total hip replacement. Bone Joint J 95(3):343–347. https://doi.org/10.1302/0301-620X.95B3.31098
Waterson HB, Whitehouse MR, Greidanus NV, Garbuz DS, Masri BA, Duncan CP (2018) Revision for adverse local tissue reaction following metal-on-polyethylene total hip arthroplasty is associated with a high risk of early major complications. Bone Joint J 100(6):720–724. https://doi.org/10.1302/0301-620X.100B6.BJJ-2017-1466.R1
Brown JM, Mistry JB, Cherian JJ, Elmallah RK, Chughtai M, Harwin SF, Mont MA (2016) Femoral component revision of total hip arthroplasty. Orthopedics 39(6):e1129–e1139. https://doi.org/10.3928/01477447-20160819-06
Carlson BC, Bryan AJ, Carrillo-Villamizar NT, Sierra RJ (2017) The utility of metal Ion trends in predicting revision in metal-on-metal total hip arthroplasty. J Arthroplast 32(9):S214–S219. https://doi.org/10.1016/j.arth.2017.02.031
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The authors declare that they have no conflict of interest.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Level of evidence: IV.
Rights and permissions
About this article
Cite this article
Bansal, T., Aggarwal, S., Dhillon, M.S. et al. Gross trunnion failure in metal on polyethylene total hip arthroplasty—a systematic review of literature. International Orthopaedics (SICOT) 44, 609–621 (2020). https://doi.org/10.1007/s00264-019-04474-z
Received:
Accepted:
Published:
Issue Date: