当前位置: X-MOL 学术Instrum. Exp. Tech. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
The Energy Characteristics of a Multichannel Scintillation Spectrometer
Instruments and Experimental Techniques ( IF 0.6 ) Pub Date : 2021-09-14 , DOI: 10.1134/s0020441221050018
V. I. Alekseev 1 , V. A. Baskov 1 , V. A. Dronov 1 , A. I. L’vov 1 , A. V. Kolzov 1 , V. V. Polyansky 1 , S. S. Sidorin 1 , E. A. Varfolomeeva 2 , Yu. F. Krechetov 3
Affiliation  

Abstract—

A multichannel scintillation spectrometer with a thickness of 2.5X0 has been calibrated on a secondary electron beam with energies of 23−280 MeV at the Pakhra accelerator of the Lebedev Physical Institute. The relative energy resolution of the spectrometer is determined by the electron energy and the spectrometer thickness. The best relative energy resolution of the spectrometer is attained at an electron energy of ~90 MeV; its values are δ = 13 and 9.4% at spectrometer thicknesses of 1.0X0 and 2.5X0, respectively.



中文翻译:

多道闪烁光谱仪的能量特性

摘要-

一台厚度为 2.5X 0的多通道闪烁光谱仪已在列别杰夫物理研究所的 Pakhra 加速器上用能量为 23-280 MeV 的二次电子束校准。光谱仪的相对能量分辨率由电子能量和光谱仪厚度决定。光谱仪的最佳相对能量分辨率是在 ~90 MeV 的电子能量下获得的;在光谱仪厚度为 1.0X 0和 2.5X 0 时,其值分别为 δ = 13 和 9.4% 。

更新日期:2021-09-15
down
wechat
bug