当前位置: X-MOL 学术Pharm. Chem. J. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Validated RP-HPLC Method Development for Gentamicin Microanalysis in Pharmaceutical Dosage Forms, Urine and Blood
Pharmaceutical Chemistry Journal ( IF 0.9 ) Pub Date : 2021-08-14 , DOI: 10.1007/s11094-021-02451-8
M. L. Maheshwari 1, 2 , N. Memon 1 , A. A. Memon 1 , M. Y. Khuhawer 3 , A. H. Memon 4
Affiliation  

Gentamicin is an important member of the aminoglycoside class of antibiotics having broader spectrum with extend to ototoxicity and nephrotoxcity. A reliable and accurate high performance liquid chromatography (HPLC) procedure was established using Agilent Zorbax Eclipse plus phenyl hexyl C18 (4.6 × 250 mm, 5 μm) column and UV detection at 330 nm. Recovery of gentamicin from biological and pharmaceutical samples was more than 95-98% after spiking the drug. The sensitivity of method is predicted by calculating a lower level of the limit of detection and lower limit of quantification (0.286 and 0.867 μg/mL, respectively. The procedure was used to quantify gentamicin in biological fluids and pharmaceuticals dosage form.



中文翻译:

用于药物剂型、尿液和血液中庆大霉素微量分析的经验证的 RP-HPLC 方法开发

庆大霉素是氨基糖苷类抗生素的重要成员,具有更广的谱,可扩展至耳毒性和肾毒性。使用 Agilent Zorbax Eclipse plus 苯基己基 C18(4.6 × 250 mm,5 μm)色谱柱和 330 nm 紫外检测建立了可靠且准确的高效液相色谱 (HPLC) 程序。加标药物后,生物和药物样品中庆大霉素的回收率超过 95-98%。通过计算检测下限和定量下限(分别为 0.286 和 0.867 μg/mL)来预测方法的灵敏度。该程序用于定量生物体液和药物剂型中的庆大霉素。