Journal of Radioanalytical and Nuclear Chemistry ( IF 1.6 ) Pub Date : 2021-08-15 , DOI: 10.1007/s10967-021-07932-4 M. J. Han 1, 2 , S. H. Hwang 1, 2 , D. H. Heo 1 , B. C. Kim 1 , B. J. Kim 1 , J. M. Lee 1, 2 , K. B. Lee 1, 2 , Y. G. Seon 3, 4
A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2).