当前位置: X-MOL 学术J. Radioanal. Nucl. Chem. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Development of a mobile radon calibration system at KRISS
Journal of Radioanalytical and Nuclear Chemistry ( IF 1.6 ) Pub Date : 2021-08-15 , DOI: 10.1007/s10967-021-07932-4
M. J. Han 1, 2 , S. H. Hwang 1, 2 , D. H. Heo 1 , B. C. Kim 1 , B. J. Kim 1 , J. M. Lee 1, 2 , K. B. Lee 1, 2 , Y. G. Seon 3, 4
Affiliation  

A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2).



中文翻译:

在 KRISS 开发移动氡校准系统

韩国标准与科学研究院 (KRISS)开发了一种移动式氡气校准系统,该系统具有高灵敏度氡气探测器、可移动校准室和226 Ra 固体源。高灵敏度氡探测器基于静电收集机制。校准室的体积为 (1003.8 ± 6.0) L。移动氡校准系统使用来自 KRISS 氡主系统的222 Rn CRM进行校准。灵敏度被确定为(231.8±16.1)的cpm /(KBQ /米3),用于218宝和(254.9±17.8)的cpm /(KBQ /米3),用于214宝,分别为(ķ  = 2)。

更新日期:2021-08-19
down
wechat
bug