A protocol for accurate radiochromic film dosimetry using Radiochromic.com.
Radiology and Oncology ( IF 2.4 ) Pub Date : 2021-08-10 , DOI: 10.2478/raon-2021-0034
Ignasi Méndez 1 , Juan José Rovira-Escutia 2 , Bozidar Casar 1
Affiliation  

BACKGROUND Radiochromic films have many applications in radiology and radiation therapy. Generally, the dosimetry system for radiochromic film dosimetry is composed of radiochromic films, flatbed scanner, and film analysis software. The purpose of this work is to present the effectiveness of a protocol for accurate radiochromic film dosimetry using Radiochromic.com as software for film analysis. MATERIALS AND METHODS Procedures for image acquisition, lot calibration, and dose calculation are explained and analyzed. Radiochromic.com enables state-of-the-art models and corrections for radiochromic film dosimetry, such as the Multigaussian model for multichannel film dosimetry, and lateral, inter-scan, and re-calibration corrections of the response. RESULTS The protocol presented here provides accurate dose results by mitigating the sources of uncertainty that affect radiochromic film dosimetry. CONCLUSIONS Appropriate procedures for film and scanner handling in combination with Radiochromic.com as software for film analysis make easy and accurate radiochromic film dosimetry feasible.

中文翻译:

使用 Radiochromic.com 进行精确放射变色胶片剂量测定的协议。

发明背景放射变色膜在放射学和放射治疗中有许多应用。通常,放射变色胶片剂量测定的剂量系统由放射变色胶片、平板扫描仪和胶片分析软件组成。这项工作的目的是展示使用 Radiochromic.com 作为胶片分析软件的准确放射变色胶片剂量测定协议的有效性。材料和方法 对图像采集、批次校准和剂量计算的程序进行了解释和分析。Radiochromic.com 支持放射变色胶片剂量测定的最先进模型和校正,例如用于多通道胶片剂量测定的多高斯模型,以及响应的横向、扫描间和重新校准校正。结果 这里介绍的协议通过减轻影响放射变色胶片剂量测定的不确定性来源提供准确的剂量结果。结论 适当的胶片和扫描仪处理程序与 Radiochromic.com 作为胶片分析软件相结合,使放射变色胶片剂量测定变得简单而准确。
更新日期:2021-08-10
down
wechat
bug