Determination of the optimal range for virtual monoenergetic images in dual-energy CT based on physical quality parameters
Medical Physics ( IF 3.8 ) Pub Date : 2021-07-21 , DOI: 10.1002/mp.15120
Enric Fernandez-Velilla Cepria 1, 2 , Miguel Ángel González-Ballester 3, 4 , Jaume Quera Jordana 1, 2, 5 , Oscar Pera 1, 2 , Xavier Sanz Latiesas 1, 2, 5 , Palmira Foro Arnalot 1, 2, 5 , Ismael Membrive Conejo 1, 2 , Nuria Rodriguez de Dios 1, 2, 5 , Anna Reig Castillejo 1, 2 , Manuel Algara Lopez 1, 2, 6
Affiliation  

Virtual monoenergetic images (VMI) obtained from Dual-Energy Computed Tomography (DECT) with iodinated contrast are used in radiotherapy of the Head and Neck to improve the delineation of target volumes and organs at-risk (OAR). The energies used to vary from 40 to 70 keV, but noise at low keV and the use of Single Energy CT (SECT) at low kVp settings may shrink this interval. There is no guide about how to find out the optimal range where VMI has a significant improvement related to SECT images. Our study proposes a procedure to determine this optimal range, based on common image quality parameters, and establishes this range in a Siemens Somatom Confidence and a Head and Neck protocol.