An Improved Reflection Colorimeter Integrated with a Coaxial Optical-fiber Cable for Highly Sensitive Solid-phase Colorimetry Using a Membrane Filter
Analytical Sciences ( IF 1.6 ) Pub Date : 2021-07-10 , DOI: 10.2116/analsci.20n033
Hitoshi MIZUGUCHI 1 , Yusuke TANIGUCHI 1 , Suherman SUHERMAN 2 , Kinichi MORITA 3 , Toshio TAKAYANAGI 1 , Yasutada SUZUKI 4
Affiliation  

Highly sensitive solid-phase colorimetry for nickel ion was demonstrated using an improved reflection colorimeter equipped with a coaxial optical-fiber cable. The nickel complex with α-furil dioxime was collected on a small-size membrane filter embedded in a disposable syringe filter unit. The leading edge of the optical-fiber cable was connected to the syringe filter unit via a Luer-lock fitting, and the color intensity of the sample on the filter was evaluated accurately. The detection limit was 0.8 ng in 2.5 mL of the complex solution (0.3 μg L−1). This improved configuration is applicable to highly sensitive on-site analysis without expensive instruments nor high laboratory skills.