Ultrafast LC-MS/MS analysis of 5-hydroxyindoleacetic acid (5-HIAA) in serum
Scandinavian Journal of Clinical and Laboratory Investigation ( IF 2.1 ) Pub Date : 2021-06-08 , DOI: 10.1080/00365513.2021.1930141
Joakim Oxelbark 1 , Annicka Lövenhamn 2
Affiliation  

Abstract

A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50–100 nmol/L) to 99.7% at high levels (500 nmol/L). A previously reported reference interval of 35–123 nmol/L was confirmed as valid based on analysis of 40 samples from voluntary blood donors.



中文翻译:

血清中 5-羟基吲哚乙酸 (5-HIAA) 的超快 LC-MS/MS 分析

摘要

已开发出一种非常快速简便的 5-HIAA(5-羟基吲哚乙酸)LC-MS/MS 分析方法。该方法经过充分验证,并证明在临床环境中运行良好。参考上限 123 nmol/L 的精密度为 3.3% CV。准确度范围从低浓度 (50–100 nmol/L) 的 96% 到高浓度 (500 nmol/L) 的 99.7%。根据对来自自愿献血者的 40 个样本的分析,确认了先前报告的 35-123 nmol/L 参考区间是有效的。