A Novel Flow-injection Rhodium Nanoparticles Modified Phosphate Biosensor and its Operation in Artificial Urine
Electroanalysis ( IF 3 ) Pub Date : 2021-06-07 , DOI: 10.1002/elan.202100154
Muhammet Kilic 1
Affiliation  

A biosensor for phosphate determination with the flow-injection system was developed using rhodium nanoparticles modified Poly(pyrrole-co-[1-(2-aminophenyl) pyrrole])/pyruvate oxidase. The biosensor showed a very wide linearity up to 70 mM phosphate concentration compared to previously reports, response time of 4 s., operational stability with a relative standard deviation of 0.009 % and accuracy of 99.4 %±0.949 at a flow rate of 2.0 Ml min.−1 at exactly −0.68 V. Detection limit were calculated to be 21±0.001 μM by preserving 81.1 % of its initial response at the end of 16th days. Artificial urine was analyzed without dilution to investigate biosensor performance.