Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients
Scandinavian Journal of Clinical and Laboratory Investigation ( IF 2.1 ) Pub Date : 2021-05-31 , DOI: 10.1080/00365513.2021.1930142
Fiamma Balboni 1 , Alessandro Terreni 2 , Marco Gallo 3 , Francesco Guzzi 3 , Chiara Caparrini 4 , Stefania Burbui 5 , Marco Vezzosi 1 , Silvia Galora 1 , Giulia Lori 1 , Giuseppe Lippi 6
Affiliation  

Abstract

GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3) levels. The aim of this work was to evaluate the clinical performance of HCO3 assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3 levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3 of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3 quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland–Altman analysis of HCO3 levels evidenced a bias (µ) of −4.9 (95% CI, −5.2 to −4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3 values from pH and pCO2 using the Henderson Hasselbach equation, µ=–0.07 (95%CI, −0.19 to 0.05) mmol/L (p = .24). In conclusion, our results show a remarkable difference between the HCO3 values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3 shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device.



中文翻译:

使用新的 gem Premier chemstat 对血液透析患者进行直接碳酸氢盐测量的分析评估

摘要

GEM Premier ChemSTAT 是一款全血分析仪,旨在提供快速的基本代谢面板,包括肌酐和血尿素氮,具有提供测量碳酸氢盐 (HCO 3 ) 水平的独特特性。这项工作的目的是评估 HCO 3的临床性能-在现实血液透析环境中使用该分析仪进行评估。计算了不同 HCO 3水平不精密度,并与 Gem Premier 4000 分析仪进行了分析比较。GEM Premier ChemSTAT 显示 HCO 3的不精确性和偏差(与 GEM Premier 4000 相比) 在 20.8 mmol/L 时为 0.4% 和 37.3%,在 16.4 mmol/L 时为 1.2% 和 25.6%,在 11.5 mmol/L 时分别为 2.1% 和 11.6%,使用三个水平的 HCO 3 质量控制样品 ChemSTAT System Evaluator . 在与血液透析环境中的 GEM Premier 4000 直接比较时,HCO 3 水平的 Bland-Altman 分析表明偏差 ( µ ) 为 -4.9(95% CI,-5.2 至 -4.7)mmol/L。通过使用 Henderson Hasselbach 方程重新计算 GEM ChemSTAT 预期的 HCO 3 来自 pH 和 pCO 2的值,可以减弱这种差异, µ =–0.07 (95%CI, -0.19 至 0.05) mmol/L ( p = .24)。总之,我们的结果显示 HCO 3 (由 GEM ChemSTAT 或 GEM 4000 报告的值)之间存在显着差异。因此应根据我们的研究结果定义GEM ChemSTAT HCO 3的新参考值。我们建议在扩大该设备的临床使用范围之前,应进一步调查和重新评估参考范围。

更新日期:2021-05-31
down
wechat
bug