Verification of the delivered patient radiation dose for non-coplanar beam therapy
Journal of Applied Clinical Medical Physics ( IF 2.1 ) Pub Date : 2021-05-22 , DOI: 10.1002/acm2.13280
Ivan Kutuzov 1, 2 , Timothy Van Beek 2 , Boyd M C McCurdy 1, 2, 3
Affiliation  

There is an increased interest in using non-coplanar beams for radiotherapy, including SBRT and SRS. This approach can significantly reduce doses to organs-at-risk, however, it requires stringent quality assurance, especially when a dynamic treatment couch is used. In this work, new functionality that allows using non-coplanar beam arrangements in addition to conventional coplanar beams was added and validated to the previously developed in vivo dose verification system.

中文翻译:

验证用于非共面射束治疗的患者辐射剂量

使用非共面光束进行放射治疗的兴趣越来越大,包括 SBRT 和 SRS。这种方法可以显着降低危及器官的剂量,但是,它需要严格的质量保证,尤其是在使用动态治疗床时。在这项工作中,除了传统的共面光束之外,还允许使用非共面光束布置的新功能被添加并验证到先前开发的体内剂量验证系统。