Comparison of Actual Performance in the Flow and Fraction of Inspired O2 among Different High-Flow Nasal Cannula Devices: A Bench Study
Canadian Respiratory Journal ( IF 2.2 ) Pub Date : 2021-05-04 , DOI: 10.1155/2021/6638048
Yuyan Zhou 1 , Zhong Ni 1 , Yuenan Ni 1 , Binmiao Liang 1 , Zongan Liang 1
Affiliation  

Background. High-flow nasal cannula (HFNC) oxygen therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure and many other clinical conditions. HFNC devices produced by different manufacturers may have varied performance. Whether there is a difference in these devices and the extent of the differences in performance remain unknown. Methods. Four HFNC devices (AIRVO 2, TNI softFlow 50, HUMID-BH, and OH-70C) and a ventilator with an HFNC module (bellavista 1000) were evaluated. The flow was set at 20, 25, 30, 35, 40, 45, 50, 60, 70, and 80 L/min, and the FiO2 was set at 21%, 26%, 30%, 35%, 40%, 45%, 50%, 60%, 70%, 80%, and 90%. Then, one side of the cannulas was clipped to simulate the compression, bending, or blocking of the nasal cannulas. The flow and FiO2 of the delivered gas were recorded and compared among settings and devices. Results. The actual-flow and actual-FiO2 delivered by different settings and devices varied. AIRVO 2 had superior performance in flow and FiO2 accuracy. bellavista 1000 and OH-70C had good performance in the accuracy of actual-flows and actual-FiO2, respectively. bellavista 1000 and HUMID-BH had a larger flow range from 10 to 80 L/min, but only bellavista 1000 could provide a stable flow with an excessive resistance up to 60 L/min. TNI softFlow 50 had the best flow compensation and could provide sufficient flow with excessive resistance at 20–50 L/min. Conclusions. The variation in flow, FiO2 settings, and devices could influence the actual-flow and actual-FiO2 delivered. AIRVO 2 and OH-70C showed better FiO2 accuracy. TNI softFlow 50, bellavista 1000, and HUMID-BH could lower the risk of insufficient flow support due to accidental compression or blocking of the cannulas. In addition, ventilators with HFNC modules provided comparable flow and FiO2 and could be an alternative to standalone HFNC devices.

中文翻译:

不同高流量鼻插管设备中吸入氧气的流量和分数的实际性能比较:基准研究

背景。已建议将高流量鼻插管(HFNC)氧气疗法用于患有急性呼吸衰竭和许多其他临床状况的冠状病毒病2019(COVID-19)患者。不同制造商生产的HFNC设备可能具有不同的性能。这些设备是否存在差异以及性能差异的程度仍然未知。方法。评估了四个HFNC设备(AIRVO 2,TNI softFlow 50,HUMID-BH和OH-70C)和带有HFNC模块的呼吸机(bellavista 1000)。流量设置为20、25、30、35、40、45、50、60、70和80 L / min,FiO 2设定为21%,26%,30%,35%,40%,45%,50%,60%,70%,80%和90%。然后,将插管的一侧剪裁以模拟鼻插管的压缩,弯曲或阻塞。记录输送气体的流量和FiO 2,并在设置和设备之间进行比较。结果。不同设置和设备提供的实际流量和实际FiO 2有所不同。AIRVO 2在流量和FiO 2精度方面具有卓越的性能。bellavista 1000和OH-70C在实际流量和实际FiO 2的准确性方面表现出色分别。bellavista 1000和HUMID-BH的流量范围较大,为10至80 L / min,但只有bellavista 1000才能提供稳定的流量,且阻力最高可达60 L / min。TNI softFlow 50具有最佳的流量补偿,并且可以在20–50 L / min的流量下提供足够的流量,但阻力过大。结论。流量,FiO 2设置和设备的变化可能会影响实际流量和实际交付的FiO 2。AIRVO 2和OH-70C显示出更好的FiO 2精度。TNI softFlow 50,bellavista 1000和HUMID-BH可以降低由于意外压缩或插管阻塞而导致流量支持不足的风险。此外,配备HFNC模块的呼吸机可提供相当的流量和FiO 2 并可以替代独立的HFNC设备。
更新日期:2021-05-04
down
wechat
bug