Biomedical Optics Express ( IF 3.4 ) Pub Date : 2021-01-27 , DOI: 10.1364/boe.412210
Giulia Maffeis 1 , Edoardo Ferocino 1 , Alberto Dalla Mora 1 , Antonio Pifferi 1, 2 , Rinaldo Cubeddu 1 , Paola Taroni 1, 2
Affiliation  

A recent upgrade of the time domain multi-wavelength optical mammograph developed by Politecnico di Milano achieved good performance in laboratory tests [Biomed. Opt. Express 9, 755 (2018). [CrossRef] ]. However, it proved unsatisfactory when in vivo measurements were finally performed. That led to a further upgrade, including the replacement of the time-to-digital converter with a new model, and the related set-up changes. The new instrument version offers improved laboratory performance (as assessed through established protocols: BIP and MEDPHOT) and good in vivo performance (extension of the scanned breast area, repeatability, consistency of estimated tissue composition with physiology). Besides introducing the new set-up and detailing its laboratory and in vivo performance, we highlight the importance of systematic in vivo testing before entering clinical trials.