The preparation of bifunctional hybrid nano-flowers and their application in the enzyme-linked immunosorbent assay for Helicobacter pylori detection
Analyst ( IF 4.2 ) Pub Date : 2020-11-07 , DOI: 10.1039/d0an01533d
Tiantian Wang 1, 2, 3, 4, 5 , Xiangguang Li 1, 2, 3, 4, 5 , Lili Chen 1, 2, 3, 4, 5 , Youhuan Zhang 1, 2, 3, 4, 5 , Yujun Zheng 1, 2, 3, 4, 5 , Linjin Yu 1, 2, 3, 4, 5 , Zhiyu Ye 5, 6, 7, 8 , Huaqian Wang 1, 2, 3, 4, 5 , Xiping Cui 1, 2, 3, 4, 5 , Suqing Zhao 1, 2, 3, 4, 5
Affiliation  

As the infection by Helicobacter pylori (H. pylori, HP) remains for a lifetime and may induce diseases such as gastric cancer, it is vital to detect and diagnose it. A new non-invasive indirect enzyme-linked immunosorbent assay (iELISA) method based on nano-flowers (NFs) is very advantageous for the sensitive detection of HP. Furthermore, the established iELISA method based on the organic–inorganic bifunctional hybrid nano-flowers including rabbit polyclonal antibody of HP labeled with peroxidase from horseradish (R-HP-Ab–HRP@Cu2+ NFs) showed linearity with HP at a concentration of 0–105 CFU mL−1 (R2 = 0.9997). Moreover, the limit of detection (LOD) reached 50 CFU mL−1, and not only was the detection sensitivity 20 times higher than that based on rabbit polyclonal antibody of HP labeled with peroxidase from horseradish (R-HP-Ab–HRP) but also the stability of R-HP-Ab–HRP in NFs was improved. In addition, the OD450 nm value was still linearly related to the concentration of HP at a range of 0–105 CFU mL−1 (R2 = 0.9952) with a LOD of 50 CFU mL−1 in an artificial saliva system. This study provided a sensitive, low-cost and convenient method for the non-invasive detection of HP.

中文翻译:

双功能杂种纳米花的制备及其在幽门螺杆菌检测中的应用

由于幽门螺杆菌H. pylori,HP)的感染会持续一生,并可能诱发胃癌等疾病,因此对其进行检测和诊断至关重要。一种新的基于纳米花(NFs)的非侵入性间接酶联免疫吸附测定(iELISA)方法对于HP的灵敏检测非常有利。此外,基于有机-无机双功能杂种纳米花(包括用辣根过氧化物酶标记的HP兔多克隆抗体(R-HP-Ab–HRP @ Cu 2+ NFs)标记的HP的兔多克隆抗体)建立的iELISA方法与HP呈线性关系。 0-10 5 CFU mL -1R 2 = 0.9997)。此外,检测限(LOD)达到50 CFU mL-1,不仅检测灵敏度比辣根过氧化物酶标记的HP兔多克隆抗体(R-HP-Ab-HRP)高20倍,而且R-HP-Ab-HRP在NFs中的稳定性得到了改善。此外,在人工唾液系统中,LOD为50 CFU mL -1时,OD 450 nm值仍与HP浓度在0-10 5 CFU mL -1 R 2 = 0.9952)范围内线性相关。这项研究为非侵入性检测HP提供了一种灵敏,低成本且方便的方法。