Evaluation of BD Phoenix and Microscan WalkAway for determination of fosfomycin susceptibility in Staphylococcus aureus.
Diagnostic Microbiology and Infectious Disease ( IF 2.9 ) Pub Date : 2020-10-17 , DOI: 10.1016/j.diagmicrobio.2020.115240
Alessandro Bondi 1 , Marco Peradotto 1 , Gabriele Bianco 1 , Davide Ghibaudo 1 , Anna Maria Barbui 1 , Cristina Costa 1 , Rossana Cavallo 1
Affiliation  

We evaluate the performance of BD Phoenix and MiscroScan WalkAway systems versus reference agar dilution method for fosfomycin susceptibility determination in 200 S.aureus clinical isolates.

Both methods exhibit ≤89.0% of Categorica Agreement and ≥63.3% of Very Major Error.

All commercial antimicrobial susceptibility systems show poor concordance with reference method.



中文翻译:

BD Phoenix 和 Microscan WalkAway 对金黄色葡萄球菌中磷霉素敏感性测定的评价。

我们评估BD凤凰MiscroScan的WalkAway系统对比基准琼脂稀释法对200个磷霉素敏感性测定性能金黄色葡萄球菌临床分离株。

两种方法均表现出 ≤ 89.0% 的 Categorica 一致性和 ≥ 63.3% 的非常重大误差。

所有商业抗微生物药敏系统与参考方法的一致性较差。