Representation and illustration of the initial parameters in GATE 8.1 monte carlo simulation of an Elekta Versa-HD linear accelerator
Journal of Radiation Research and Applied Sciences ( IF 1.7 ) Pub Date : 2020-09-30 , DOI: 10.1080/16878507.2020.1820271
Fouad A. Abolaban 1 , Eslam M. Taha 1
Affiliation  

ABSTRACT

The purpose of this study is to validate GATE 8.1 Radiotherapy Monte Carlo (MC) simulation of a 6 MV Elekta Versa-HD linear accelerator at King Abdulaziz University Hospital (KAUH). The simulation data using GATE 8.1 was benchmarked against measured data at KAUH. The simulation comprises treatment head, Multi-Leaf Collimators (MLCs), and homogeneous water phantom, whereas the measured data was performed using MP3 phantom and PTW 31,010, 0.125 cm3 Semiflex Chamber. Dose depth distribution and dose profiles were carried out at three different field sizes 10x10, 20x20, and 30 × 30 cm2. The calculated TPR20,10 for the simulated data was found to be 0.658, 1.1% less than the measured TPR20,10 which was found to be 0.666. The obtained results indicate good agreement between the simulated and measured data, where Gamma Index 3%/3 mm criteria reached values of 97% and 90% for relative dose and dose profiles, respectively.



中文翻译:

Elekta Versa-HD线性加速器的GATE 8.1蒙特卡罗模拟中的初始参数表示和图示

摘要

这项研究的目的是验证阿卜杜勒阿齐兹国王大学医院(KAUH)的6 MV Elekta Versa-HD线性加速器的GATE 8.1放射治疗蒙特卡罗(MC)模拟。使用GATE 8.1的仿真数据与KAUH的测量数据进行了基准比较。该模拟包括处理头,多叶准直器(MLC)和均质水体模,而测量数据是使用MP3体模和PTW 31,010,0.125 cm 3 Semiflex Chamber进行的。剂量深度分布和剂量分布在三种不同的场大小10x10、20x20和30×30 cm 2下进行。发现仿真数据的TPR 20,10为0.658,比实测TPR 20,10低1.1%被发现是0.666。获得的结果表明,模拟数据与测量数据之间具有良好的一致性,其中相对剂量和剂量曲线的Gamma指数3%/ 3 mm标准分别达到97%和90%的值。

更新日期:2020-09-30
down
wechat
bug