Imaging of two samples with a single transmit/receive channel using coupled ceramic resonators for MR microscopy at 17.2 T.
NMR in Biomedicine ( IF 2.9 ) Pub Date : 2020-08-31 , DOI: 10.1002/nbm.4397
Marine A C Moussu 1, 2 , Stanislav B Glybovski 3 , Redha Abdeddaim 2 , Christophe Craeye 4 , Stefan Enoch 2 , Denis Tihon 5 , Sergej Kurdjumov 3 , Marc Dubois 2 , Elodie Georget 1 , Andrew G Webb 6 , Pavel Belov 3 , Luisa Ciobanu 7
Affiliation  

In this paper we address the possibility to perform imaging of two samples within the same acquisition time using coupled ceramic resonators and one transmit/receive channel. We theoretically and experimentally compare the operation of our ceramic dual‐resonator probe with a wire‐wound solenoid probe, which is the standard probe used in ultrahigh‐field magnetic resonance microscopy. We show that due to the low‐loss ceramics used to fabricate the resonators, and a favorable distribution of the electric field within the conducting sample, a dual probe, which contains two samples, achieves an SNR enhancement by a factor close to the square root of 2 compared with a solenoid optimized for one sample.