Dosimetry formalism and calibration procedure for electronic brachytherapy sources in terms of absorbed dose to water
Physics in Medicine & Biology ( IF 3.5 ) Pub Date : 2020-07-26 , DOI: 10.1088/1361-6560/ab9772
A Abudra’a 1 , B Chauvenet 1 , J Gouriou 1 , J Plagnard 1 , R Itti 2 , I Aubineau-Lanice 1
Affiliation  

The LNE-LNHB has developed a methodology to standardize electronic brachytherapy sources in terms of absorbed dose to water. It is based on the measurement of the air-kerma rate at a given distance from the source and the Monte Carlo calculation of a conversion factor. This factor converts the air-kerma in measurement conditions into absorbed dose to water at a 1 cm reference depth in a water phantom. As a first application, the method was used to calibrate a Zeiss INTRABEAM system equipped with its 4 cm diameter spherical applicator. The absorbed-dose rate value obtained in the current study was found significantly higher than that provided by the manufacturer in line with the observations already reported by a few other teams.