Final report on key comparison EURAMET.AUV.V-K2
Metrologia ( IF 2.4 ) Pub Date : 2020-07-26 , DOI: 10.1088/0026-1394/57/1a/09003
Joanna Kolasa 1 , Daniela Virovska 2
Affiliation  

This final report describes the results obtained in a key comparison of primary vibration calibration facilities. Two piezoelectric standard accelerometers (one single-ended as well one back-to-back type) were calibrated first at Central Office of Measures (GUM), acting as pilot and linking laboratory, and Bulgarian Institute for Metrology (BIM). The results of the BIM obtained in this comparison were linked to the CCAUV.V-K2 key comparison through the joint participation of the GUM. The degrees of equivalence were computed for BIM with respect to the CCAUV.V-K2 key comparison reference value. == Main text To reach the main text of this paper, click on Final Report [https://www.bipm.org/utils/common/pdf/final_reports/AUV/V-K2/EURAMET.AUV.V-K2.pdf] . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ [http://kcdb.bipm.org/] . The final report has been peer-reviewed and approved ...

中文翻译:

关于关键比较EURAMET.AUV.V-K2的最终报告

这份最终报告描述了在主要振动校准设施的关键比较中获得的结果。首先在作为测试和链接实验室的中央测量办公室(GUM)和保加利亚计量学会(BIM)校准了两个压电标准加速度计(一个单端以及一个背对背型)。通过GUM的联合参与,在此比较中获得的BIM结果与CCAUV.V-K2关键比较相关。计算BIM相对于CCAUV.V-K2键比较参考值的等效度。==正文要访问本文的正文,请单击“最终报告[https://www.bipm.org/utils/common/pdf/final_reports/AUV/V-K2/EURAMET.AUV.V-K2。 pdf]。请注意,此文本显示在BIPM密钥比较数据库kcdb的附录B中。bipm.org/ [http://kcdb.bipm.org/]。最终报告已经同行评审并通过...