Instrument and method for measuring the dose rates in 60Co γ rays environment with a 2K×2K CMOS image sensor imaging system
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment ( IF 1.4 ) Pub Date : 2020-07-22 , DOI: 10.1016/j.nima.2020.164451
Yuanyuan Xue , Zujun Wang , Qianli Jiao , Minbo Liu , Hao Ning , Rui Xu , Zhibin Yao , Baoping He , Wuying Ma

The Instrument and method for measuring the dose rates in 60Co γ rays environment with a 2K×2K CMOS image sensor (CIS) imaging system are presented. The experiments were carried out at 60Co γ rays facility. The output signal increase, the frequency of effective spike were measured by the CIS imaging system under the γ rays environment at various dose rates. The dose rates were also measured by the PTW-UNIDOS. Besides, the relationship between mean output signal increase, frequency of effective spike, and dose rates are analyzed and used to measure the dose rates. The dose rates measured by our measurement system are in good agreement with the PTW-UNIDOS results. Usually, using the frequency of effective spikes to measure the dose rates would be more efficient but it is also more complex. The measuring ranges, the life span of the CIS imaging system are also discussed. This work can provide a great way to measure dose rates in 60Co γ rays environment.



中文翻译:

测量60 Co中剂量率的仪器和方法γ 2K的光线环境×2K CMOS图像传感器成像系统

测量60 Co中剂量率的仪器和方法γ 2K的光线环境×介绍了2K CMOS图像传感器(CIS)成像系统。实验在60 Coγ射线设施。输出信号增加,有效尖峰的频率由CIS成像系统在γ各种剂量率的射线环境。剂量率也通过PTW-UNIDOS测量。此外,分析了平均输出信号增加,有效尖峰频率和剂量率之间的关系,并将其用于测量剂量率。我们的测量系统测量的剂量率与PTW-UNIDOS结果非常吻合。通常,使用有效尖峰的频率来测量剂量率会更有效,但也会更加复杂。还讨论了CIS成像系统的测量范围,寿命。这项工作可以提供一种测量60 Co中剂量率的好方法γ 光线环境。

更新日期:2020-07-22
down
wechat
bug