Physics in Medicine & Biology ( IF 3.5 ) Pub Date : 2020-07-07 , DOI: 10.1088/1361-6560/ab8953
Masoud Elhamiasl , Johan Nuyts

In computed tomography (CT)-imaging an optimal compromise between the radiation burden and the image quality for the imaging task is needed. Lower-dose CT is desirable, however, lowering the dose results in a lower signal-to-noise ratio and therefore in a reduced image quality. In this research, we aim to develop a tool to simulate lower-dose scans from an existing standard-dose scan. The main application of this tool is to determine the lowest possible radiation dose that still produces sufficient clinical information. The x-ray tube current reduction is modeled by estimating the noise equivalent number of photons in the high exposure scan and applying a thinning technique to reduce that number. The proposed method accounts for the bowtie filter, for the electronic system noise, for the noise correlation between neighboring detector elements, for the beam hardening effect, and for the non-linear smoothing filter in very low-dose scans. Several phantom studies with different acq...