Collimator and energy window optimization for practical imaging protocol and quantification of Yttrium-90 bremsstrahlung spect/ct: A phantom study.
Radiation Physics and Chemistry ( IF 2.9 ) Pub Date : 2021-01-01 , DOI: 10.1016/j.radphyschem.2020.109080
Ong Saw Huey , Yong Jie See , Shahira Nabila , Heng Siew Ping , Ida Suzanah

Abstract: Yttrium-90 (90Y) bremsstrahlung imaging is very complex which increases the demands on the imaging protocol and reconstruction. We evaluate the choice of collimator and energy windows (EWs) settings. All images were acquired using low-energy high-resolution (LEHR) collimator, medium-energy general-purpose (MEGP) and high-energy general-purpose (HEGP) in different energy windows (EWs) settings, ranging from 60 keV to 300 keV. The EW based scatter-correction technique was adopted by the authors. The planar images of a customised cylindrical acrylic phantom with a90Y chloride solution vial positioned in the center of phantom were assessed based on image quality, signal-to-background ratio (SBR) and coefficient of variation (CV). HEGP has the highest SBR followed by MEGP and LEHR whereas LEHR has the lowest CV followed by MEGP and HEGP. EW with two energy windows always has highest SBR followed by three energy windows and one energy windows. In term of image quality, MEGP with 2 or 3 energy windows have the images with best spatial resolution compared to septa effect in HEGP images and grainy images with poor resolution in LEHR images. MEGP collimator with 2 energy windows gave good results in term of good spatial resolution and moderate SBR and CV. The proposed imaging protocol in planar imaging is readily to be used to tailor 90Y SPECT/CT acquisition, attenuation correction (AC), scatter correction (SC) and reconstruction protocols for improved 90Y imaging in our hospital.

中文翻译:

用于 Yttrium-90 轫致辐射光谱/ct 的实用成像协议和量化的准直器和能量窗口优化:一项幻像研究。

摘要:钇 90 (90Y) 轫致辐射成像非常复杂,这增加了对成像协议和重建的要求。我们评估了准直器和能量窗 (EW) 设置的选择。所有图像均使用低能高分辨率 (LEHR) 准直器、中能通用 (MEGP) 和高能通用 (HEGP) 在不同能量窗口 (EW) 设置下获取,范围从 60 keV 到 300 keV。作者采用了基于 EW 的散射校正技术。根据图像质量、信号背景比 (SBR) 和变异系数 (CV) 评估定制的圆柱形丙烯酸模型的平面图像,该模型具有位于模型中心的 90 Y 氯化物溶液瓶。HEGP 的 SBR 最高,其次是 MEGP 和 LEHR,而 LEHR 的 CV 最低,其次是 MEGP 和 HEGP。具有两个能量窗的 EW 总是具有最高的 SBR,其次是三个能量窗和一个能量窗。在图像质量方面,与 HEGP 图像中的隔膜效应和 LEHR 图像中分辨率较差的颗粒图像相比,具有 2 或 3 个能量窗口的 MEGP 具有最佳空间分辨率的图像。具有 2 个能量窗的 MEGP 准直器在良好的空间分辨率和适中的 SBR 和 CV 方面给出了良好的结果。所提出的平面成像成像协议很容易用于定制 90Y SPECT/CT 采集、衰减校正 (AC)、散射校正 (SC) 和重建协议,以改进我们医院的 90Y 成像。与 HEGP 图像中的隔膜效应和 LEHR 图像中分辨率较差的颗粒状图像相比,具有 2 或 3 个能量窗口的 MEGP 具有最佳空间分辨率的图像。具有 2 个能量窗的 MEGP 准直器在良好的空间分辨率和适中的 SBR 和 CV 方面给出了良好的结果。所提出的平面成像成像协议很容易用于定制 90Y SPECT/CT 采集、衰减校正 (AC)、散射校正 (SC) 和重建协议,以改进我们医院的 90Y 成像。与 HEGP 图像中的隔膜效应和 LEHR 图像中分辨率较差的颗粒状图像相比,具有 2 或 3 个能量窗口的 MEGP 具有最佳空间分辨率的图像。具有 2 个能量窗的 MEGP 准直器在良好的空间分辨率和适中的 SBR 和 CV 方面给出了良好的结果。所提出的平面成像成像协议很容易用于定制 90Y SPECT/CT 采集、衰减校正 (AC)、散射校正 (SC) 和重建协议,以改进我们医院的 90Y 成像。
更新日期:2021-01-01
down
wechat
bug