当前位置: X-MOL 学术J. Solid State Electr. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Connection of CVs and impedance spectra of reversible redox systems, as used for the validation of a dynamic electrochemical impedance spectrum measurement system
Journal of Solid State Electrochemistry ( IF 2.5 ) Pub Date : 2020-05-31 , DOI: 10.1007/s10008-020-04661-8
Tamás Pajkossy , Gábor Mészáros

With the purpose of fast characterization of electrode reactions, a dynamic electrochemical impedance spectrum (dEIS) measurement system has been assembled which permits the continuous collection of audio-frequency impedance spectra while performing cyclic voltammetry measurements with the usual scan rates of up to 200 mV/s. The performance of this system was tested by analyzing the CV curves and impedance spectra taken simultaneously in ferro-/ferricyanide containing aqueous solutions yielding an experimental demonstration of the connection of the semi-integrated reversible voltammograms and the Warburg coefficients.

Graphical abstract



中文翻译:

CV和可逆氧化还原系统的阻抗谱的连接,用于验证动态电化学阻抗谱测量系统

为了快速表征电极反应,已组装了动态电化学阻抗谱(dEIS)测量系统,该系统可连续采集音频阻抗谱,同时以高达200 mV / V的常规扫描速率执行循环伏安法测量s。通过分析在含铁氰化铁溶液中同时获得的CV曲线和阻抗谱,测试了该系统的性能,得出了半积分可逆伏安图和Warburg系数之间联系的实验证明。

图形概要

更新日期:2020-05-31
down