当前位置: X-MOL 学术IEEE T. Magn. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Magnetic field integral measurements with stretched wire and Hall probe methods
IEEE Transactions on Magnetics ( IF 2.1 ) Pub Date : 2020-05-01 , DOI: 10.1109/tmag.2020.2976031
Mona Gehlot , Saif Mohd Khan , Frederic Trillaud , G. Mishra

In this article, we analyze magnetic measurements on the U50-II undulator. The U50-II undulator is a 50 mm period length NdFeB-based pure permanent-magnet undulator prototype. It is designed with 20 periods and delivers 0.88–0.17 T in the gap from 10 to 35 mm. The field integrals are measured on a Hall probe bench, and the measurements are compared with the stretched wire result. In the stretched wire setup, the field integral is measured with an accuracy of 2 mTmm. The details of the stretch wire probe setup, measurement procedure, and accuracy with Hall probe results are discussed.

中文翻译:

使用拉伸线和霍尔探针方法进行磁场积分测量

在本文中,我们分析了 U50-II 波荡器的磁性测量。U50-II 波荡器是一个 50 毫米周期长度的 NdFeB 基纯永磁波荡器原型。它设计有 20 个周期,并在 10 到 35 毫米的间隙中提供 0.88-0.17 T。在霍尔探针台上测量场积分,并将测量结果与拉伸线结果进行比较。在拉伸线设置中,场积分的测量精度为 2 mTmm。讨论了拉伸线探头设置、测量程序和霍尔探头结果精度的详细信息。