Mobile-bearing insert used with total knee arthroplasty does not rotate on the tibial tray during a squatting activity: a cross-sectional study
Journal of Orthopaedic Surgery and Research ( IF 2.6 ) Pub Date : 2020-03-20 , DOI: 10.1186/s13018-020-1570-6
Kenji Hoshi , Goro Watanabe , Yasuo Kurose , Ryuji Tanaka , Jiro Fujii , Kazuyoshi Gamada

Total knee arthroplasty (TKA) is commonly performed around the world. Implant designs include fixed-bearing and mobile-bearing. Mobile-bearing design was developed as a rotating platform that allows axial rotation of the insert around the longitudinal axis. This phenomenon may limit full exploitation of the characteristics of the mobile-bearing insert, which may cause wearing and reduce longevity. However, there is limited knowledge on rotational behavior of the polyethylene mobile-bearing insert under weight-bearing conditions. We aimed at determining the rotational motion of each component at full extension and flexed positions during a squatting activity after TKA. This study was a cross-sectional study (level 4) involving patients with severe knee osteoarthritis scheduled to receive TKA. We examined 13 knees of 11 patients after mobile-bearing TKA (NexGen LPS-Flex, Zimmer Inc.) at 10 weeks and 1 year postoperatively. Four identical metallic beads were embedded into the insert. Wide-base squatting was chosen for analyses. Three-dimensional in vivo poses of the prostheses were created using a 3D-to-2D registration technique. During flexion, rotation of the femoral component relative to the insert (FEM/INS) and tibial component (FEM/TIB) as well as insert rotation relative to the tibial component (INS/TIB) were computed. Repeated measure 2-way ANOVA and post hoc test was used. In the fully extended position, FEM/INS was significantly smaller than INS/TIB both at 10 weeks (− 0.3° vs. 6.3°, p = .013) and 1 year (− 0.8° vs. 4.9°, p = .011), respectively. During the squatting activity, rotation motions of FEM/TIB, FEM/INS, INS/TIB were 5.7°, 5.9°, and 1.8° at 10 weeks and 6.3°, 5.5°, and 1.6° at 1 year, respectively. Rotation motion of FEM/INS was significantly greater than that of INS/TIB at both 10 weeks (p < .001) and 1 year (p < .001). The mobile-bearing insert enhances the compatibility of FEM/INS in extension; the amount of INS/TIB rotation is significantly smaller than that of FEM/INS during a squatting activity. This information will inform surgeons to take caution to perform TKA with a fixed insert in which 6.3° of rotational offset would be added to the rotational alignment at FEM/INS at full extension. UMIN-CTR, UMIN000024196. Retrospectively registered on 9 September 2016.

中文翻译:

进行全蹲式活动时,与全膝关节置换术一起使用的活动轴承插入物不会在胫骨托上旋转

全膝关节置换术(TKA)在世界范围内普遍进行。植入物设计包括固定轴承和移动轴承。移动轴承设计被开发为旋转平台,该旋转平台允许刀片围绕纵向轴线轴向旋转。这种现象可能会限制充分利用可动轴承插件的特性,这可能会导致磨损并降低使用寿命。然而,关于聚乙烯可移动轴承插入物在承重条件下的旋转行为的知识有限。我们的目标是确定在TKA后的下蹲活动期间每个组件在完全伸展和弯曲位置的旋转运动。这项研究是一项横断面研究(4级),涉及计划接受TKA的重度膝骨关节炎患者。我们在术后10周和1年时检查了11例活动性TKA(NexGen LPS-Flex,Zimmer Inc.)患者的13个膝盖。将四个相同的金属珠嵌入插入物中。选择宽底蹲式进行分析。使用3D到2D配准技术创建了假体的三维体内姿势。在屈曲过程中,计算了股骨相对于插入物(FEM / INS)和胫骨组件(FEM / TIB)的旋转以及插入物相对于胫骨组件的旋转(INS / TIB)。使用重复测量的2-方差分析和事后检验。在完全伸展的位置,FEM / INS在10周(-0.3°vs. 6.3°,p = .013)和1年(-0.8°vs. 4.9°,p = .011)时均显着小于INS / TIB。 ), 分别。在下蹲过程中,FEM / TIB,FEM / INS,INS / TIB在10周时分别为5.7°,5.9°和1.8°,在1年时分别为6.3°,5.5°和1.6°。在10周(p <.001)和1年(p <.001)时,FEM / INS的旋转运动明显大于INS / TIB。可移动轴承的插件增强了FEM / INS在扩展方面的兼容性;在下蹲过程中,INS / TIB旋转量明显小于FEM / INS。该信息将通知外科医生注意使用固定刀片进行TKA,其中在完全伸展时,将在FEM / INS上将6.3°的旋转偏移量添加到旋转对准中。UMIN-CTR,UMIN000024196。追溯登记于2016年9月9日。在10周(p <.001)和1年(p <.001)时,FEM / INS的旋转运动明显大于INS / TIB。可移动轴承的插件增强了FEM / INS在扩展方面的兼容性;在下蹲过程中,INS / TIB旋转量明显小于FEM / INS。该信息将通知外科医生注意使用固定刀片进行TKA,其中在完全伸展时,将在FEM / INS上将6.3°旋转偏移添加到旋转对准中。UMIN-CTR,UMIN000024196。追溯登记于2016年9月9日。在10周(p <.001)和1年(p <.001)时,FEM / INS的旋转运动均明显大于INS / TIB。可移动轴承的插件增强了FEM / INS在扩展方面的兼容性;在下蹲过程中,INS / TIB旋转量明显小于FEM / INS。该信息将通知外科医生注意使用固定刀片进行TKA,其中在完全伸展时,将在FEM / INS上将6.3°旋转偏移添加到旋转对准中。UMIN-CTR,UMIN000024196。追溯登记于2016年9月9日。该信息将通知外科医生注意使用固定刀片进行TKA,其中在完全伸展时,将在FEM / INS上将6.3°旋转偏移添加到旋转对准中。UMIN-CTR,UMIN000024196。追溯登记于2016年9月9日。该信息将通知外科医生注意使用固定刀片进行TKA,其中在完全伸展时,FEM / INS上的旋转对准将增加6.3°的旋转偏移量。UMIN-CTR,UMIN000024196。追溯登记于2016年9月9日。
更新日期:2020-04-22
down
wechat
bug