当前位置: X-MOL 学术Opt. Lett. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Real-time standoff detection of nitrogen isotopes in ammonia plumes using a swept external cavity quantum cascade laser
Optics Letters ( IF 3.6 ) Pub Date : 2018-08-16 , DOI: 10.1364/ol.43.004065
Mark C. Phillips , Brian E. Brumfield , Sivanandan S. Harilal

A swept-wavelength external cavity quantum cascade laser is used to measure time-resolved NH315 and NH314 concentrations and isotope ratios in a turbulent plume source located at 10 m standoff distance. Measurements show excellent agreement with expected values of isotope ratios for samples with natural and enriched N15 abundance. Detection sensitivity of 150 ppb*m is demonstrated for a 50 ms measurement interval, improving to 10 ppb*m with 10 s averaging, with isotopic precision as good as 0.8%.

中文翻译:

使用扫频外腔量子级联激光实时检测氨气中的氮同位素

扫描波长外腔量子级联激光器用于测量时间分辨 NH315NH314距离为10 m的湍流烟羽源中的浓度和同位素比。测量结果表明,天然和富集样品的同位素比预期值具有极好的一致性ñ15丰富。在50 ms的测量间隔内显示出150 ppb * m的检测灵敏度,平均10 s提高到10 ppb * m,同位素精度高达0.8%。
更新日期:2018-08-31