Overcoming spectral overlap via inductively coupled plasma-tandem mass spectrometry (ICP-MS/MS). A tutorial review
Journal of Analytical Atomic Spectrometry ( IF 3.4 ) Pub Date : 2017-02-17 00:00:00 , DOI: 10.1039/c7ja00010c
Eduardo Bolea-Fernandez 1, 2, 3, 4, 5 , Lieve Balcaen 1, 2, 3, 4, 5 , Martín Resano 2, 6, 7, 8, 9 , Frank Vanhaecke 1, 2, 3, 4, 5
Affiliation  

This work reviews the operating principles of ICP-tandem mass spectrometry (ICP-MS/MS) and the key applications reported on since the introduction of the technique in 2012. The main differences between single quadrupole ICP-MS and ICP-MS/MS are elucidated and the tools available for method development are addressed. Examples of real-life applications are given to demonstrate the capabilities of this novel setup in the context of elemental, speciation and isotopic analysis. Attention has been paid to the different approaches (on-mass and mass-shift) allowing interference-free conditions to be obtained.

中文翻译:

通过电感耦合等离子体串联质谱(ICP-MS / MS) 克服光谱重叠。教程复习

这项工作回顾了ICP串联质谱(ICP-MS / MS)的工作原理以及自2012年引入该技术以来所报告的关键应用。单四极杆ICP-MS和ICP-MS / MS之间的主要区别是阐明并解决了可用于方法开发的工具。给出了实际应用的示例,以在元素分析,形态分析和同位素分析的背景下演示这种新颖设置的功能。注意力已经转移到允许获得无干扰条件的不同方法(质量转移和质量转移)上。